Development of design methods for a centrifugal blood pump with a fluid dynamic approach: results in hemolysis tests.

T Masuzawa1, T Tsukiya, S Endo, E Tatsumi, Y Taenaka, H Takano, T Yamane, M Nishida, B Asztalos, Y Miyazoe, K Ito, T Sawairi, Y Konishi.   

Abstract

The purpose of this study was to examine the relationship between local flow conditions and the hemolysis level by integrating hemolysis tests, flow visualization, and computational fluid dynamics to establish practical design criteria for centrifugal blood pumps with lower levels of hemolysis. The Nikkiso centrifugal blood pump was used as a standard model, and pumps with different values of 3 geometrical parameters were tested. The studied parameters were the radial gap between the outer edge of the impeller vane and the casing wall, the position of the outlet port, and the discharge angle of the impeller vane. The effect of a narrow radial gap on hemolysis was consistent with no evidence that the outlet port position or the vane discharge angle affected blood trauma in so far as the Nikkiso centrifugal blood pump was concerned. The radial gap should be considered as a design parameter of a centrifugal blood pump to reduce blood trauma.

Entities:  

Mesh:

Year:  1999        PMID: 10463503     DOI: 10.1046/j.1525-1594.1999.06417.x

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


  2 in total

1.  Estimation of changes in dynamic hydraulic force in a magnetically suspended centrifugal blood pump with transient computational fluid dynamics analysis.

Authors:  Toru Masuzawa; Akiko Ohta; Nobuatu Tanaka; Yi Qian; Tomonori Tsukiya
Journal:  J Artif Organs       Date:  2009-09-19       Impact factor: 1.731

2.  Hemolysis caused by surface roughness under shear flow.

Authors:  Osamu Maruyama; Yusuke Numata; Masahiro Nishida; Takashi Yamane; Ikuya Oshima; Yoshikazu Adachi; Toru Masuzawa
Journal:  J Artif Organs       Date:  2005       Impact factor: 1.385