| Literature DB >> 10457602 |
M J Legorburu1, R M Alonso, R M Jiménez, E Ortiz.
Abstract
A high-performance liquid chromatographic method with amperometric detection for the determination of the diuretic indapamide using a muBondapak C18 column is developed. The mobile phase consists of an acetonitrile-water mixture (45:55, 5 mM) in KH2PO4-K2HPO4 (pH 4.0). The compound is monitored at +1200 mV with an amperometric detector equipped with a glassy carbon working electrode. A liquid-liquid or solid-liquid extraction is performed prior to chromatographic analysis to avoid the interferences found in urine matrix. Percentages of recovery are 88.3 +/- 5.6 and 82.9 +/- 7.8 for liquid-liquid and solid-liquid extraction, respectively. The developed method has a linear concentration range from 25 to 315 ng/mL with a reproducibility in terms of relative standard deviation of 4% for a concentration level of 0.5 microgram/mL and a quantitation limit of 1 ng/mL. The method is applied to the determination of indapamide in tablets and urine obtained from hypertensive patients after the ingestion of Tertensif (indapamide 2.5 mg).Entities:
Mesh:
Substances:
Year: 1999 PMID: 10457602 DOI: 10.1093/chromsci/37.8.283