Thermal filament continuous thermodilution cardiac output delayed response limits its value during acute hemodynamic instability.

L F Poli de Figueiredo1, L M Malbouisson, E Y Varicoda, M J Carmona, J O Auler, M Rocha e Silva.   

Abstract

BACKGROUND: It has been suggested that measurement of continuous cardiac output (CCO) is an advancement in the management of critically ill patients. Our objective was to determine the accuracy of CCO during the rapid hemodynamic changes induced by hemorrhage and resuscitation.
METHODS: In 12 anesthetized dogs (20.2+/-0.9 kg), pulmonary artery blood flow, our "gold standard" cardiac output, was measured with an sonographic flowprobe, whereas CCO, intermittent bolus cardiac output (ICO), and mixed venous oxygen saturation were measured with a thermodilution fiberoptic pulmonary artery catheter with a thermal filament. A graded hemorrhage (20 mL/min) was produced to a mean arterial pressure of 40 mm Hg, which was maintained at this level for 30 minutes. Total shed blood volume (701+/-53 mL) was retransfused at a rate of 40 mL/min, over 30 minutes, after which a massive hemorrhage (100 mL/min) was produced over 10 minutes.
RESULTS: Hemorrhage induced significant decreases in mean arterial pressure, mixed venous oxygen saturation, and oxygen delivery, which were all restored during early resuscitation. However, CCO showed a delayed response after hemorrhage and resuscitation, compared with pulmonary blood flow, throughout the study (r = 0.549), matching only at baseline and at the end of both graded hemorrhage and resuscitation periods. There was a good correlation between ICO and pulmonary artery blood flow (r = 0.964) and no significant differences between them throughout the study.
CONCLUSION: CCO has a delayed response during acute hemodynamic changes induced by hemorrhage and resuscitation. When sudden changes in mean arterial pressure or in mixed venous oxygen saturation are detected, cardiac output must be estimated by the standard bolus thermodilution technique, not by CCO.

Entities:  

Mesh:

Year:  1999        PMID: 10452463     DOI: 10.1097/00005373-199908000-00011

Source DB:  PubMed          Journal:  J Trauma        ISSN: 0022-5282


  2 in total

1.  Tracking changes in cardiac output: methodological considerations for the validation of monitoring devices.

Authors:  Pierre Squara; Maurizio Cecconi; Andrew Rhodes; Mervyn Singer; Jean-Daniel Chiche
Journal:  Intensive Care Med       Date:  2009-07-11       Impact factor: 17.440

2.  Dynamic device properties of pulse contour cardiac output during transcatheter aortic valve implantation.

Authors:  Martin Petzoldt; Carsten Riedel; Jan Braeunig; Sebastian Haas; Matthias S Goepfert; Hendrik Treede; Stephan Baldus; Alwin E Goetz; Daniel A Reuter
Journal:  J Clin Monit Comput       Date:  2014-10-30       Impact factor: 2.502

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.