An assessment of the luminance and light field characteristics of used direct laryngoscopes.

E Crosby1, M Cleland.   

Abstract

PURPOSE: To determine luminance and light field characteristics and the effect of residual battery potential and luminance on light colour temperature in our used Macintosh #3 and #4 fibre-light (FLB) and bulb-light (BLB) laryngoscopes.
METHODS: We used a power supply to provide laryngoscopes with potentials equivalent to those most commonly measured in the handles in use in our OR. Measurements were made under controlled, constant, conditions using a Pentax digital spotmeter (luminance) or a Minolta Color III colour temperature meter (light colour). Colour measurements were made while increasing the power source potential from 2-3 volts (v) in increments of 0.1 v. Light field measurements were made with a mm increment ruler mounted on the base of the test fixture.
RESULTS: At 2.5 and 2.8 v respectively, the #3 FLB produced luminance values of 23.9 +/- 11.4 and 41.7 +/- 17.2 cd x m(-2) (mean +/- SD), and the #4 FLB produced 58.6 +/- 21.4 and 90.9 +/- 32.2 cd x m(-2). Increasing potential increased luminance values (P < 0.001) and the #4FLB produced higher luminance values (P < 0.001). BLB produced higher luminance values than did FLB across all comparisons (P < 0.001). As potentials and luminance values decreased, light temperature was reduced (P < 0.001). There were no differences in light field dimensions noted in any comparison.
CONCLUSION: Fifteen percent of the BLB did not meet the minimum luminance for laryngoscopy of 100 cd x m(-2), 92% of the FLB did not meet that same standard.

Mesh:

Year:  1999        PMID: 10451142     DOI: 10.1007/BF03013918

Source DB:  PubMed          Journal:  Can J Anaesth        ISSN: 0832-610X            Impact factor:   5.063


  2 in total

1.  A New Video Laryngoscope - An Aid in Intubation and Teaching.

Authors:  Marshal B Kaplan; Denham S Ward; George Berci
Journal:  J Educ Perioper Med       Date:  2003-01-01

2.  Videolaryngoscopes differ substantially in illumination of the oral cavity: A manikin study.

Authors:  Barbe Ma Pieters; André Aj van Zundert
Journal:  Indian J Anaesth       Date:  2016-05
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.