An accurate adjustable applicator for magnetic resonance imaging-based stereotactic procedure using the Leksell G frame.

T Kamiryo1, E R Laws.   

Abstract

OBJECTIVE: An applicator system for the Leksell G frame was constructed to enable accurate placement of the frame for stereotactic magnetic resonance imaging (MRI) and successful stereotactic surgery. The applicator prevents inaccurate placement of the fiducial box on the patient's head and prevents contact of the frame holder with the patient's shoulder while in the MRI unit. It also helps to ensure optimal positioning of desired targets within the three-dimensional stereotactic space defined by the frame.
METHODS: The applicator is made of transparent acrylic plates, which simulate the fiducial box that is attached to the frame for the preoperative stereotactic MRI study. An air cuff at the top supports the frame at any desired height and makes minute adjustments possible. Side cuffs help to keep the frame at the desired position from right to left. Indicators attached to the frame for the anterior fiducial plate prevent potential contact of the plate with the anterior posts and help avoid a poor fit caused by bending of the frame from excessive torque on the cranium fixation screws. Indicators for the MRI frame holder on the foot screws predict potential collision of the holder with the patient's shoulder before actually applying the holder on the frame. The applicator shows the range and limits of the Leksell stereotactic arc.
RESULTS: This applicator system has been used effectively in more than 89 MRI-based functional stereotactic procedures. These include pallidotomy, thalamotomy, implantation of deep brain stimulators, and implantation of depth electrodes. It has functioned well and has facilitated excellent operative results in these cases.
CONCLUSION: This simple frame applicator eliminates the need for reapplication of the stereotactic frame and additional imaging studies, thus providing successful and appropriate frame placement for stereotactic surgery.

Entities:  

Mesh:

Year:  1999        PMID: 10449088     DOI: 10.1097/00006123-199908000-00043

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  2 in total

1.  Bite-bar stabilization system for stereotactic frame application: A technical note.

Authors:  Keith Aronyk; Gregory Bowden; Angela Pickering; Kim Chandler; Samir Patel
Journal:  J Radiosurg SBRT       Date:  2020

2.  Techniques for the Application of Stereotactic Head Frames Based on a 25-Year Experience.

Authors:  Michael Safaee; John Burke; Michael W McDermott
Journal:  Cureus       Date:  2016-03-25
  2 in total