Literature DB >> 10437760

Handheld metal detector localization of ingested metallic foreign bodies: accurate in any hands?

K Seikel1, P A Primm, B J Elizondo, K L Remley.   

Abstract

OBJECTIVE: To demonstrate the accuracy of handheld metal detectors (HHMDs) for identification and localization of ingested metallic foreign bodies when used by experienced and inexperienced investigators.
DESIGN: Prospective study comparing HHMD scanning with radiography. SETTING AND PATIENTS: A consecutive sample of all eligible patients (N = 176) presenting to the emergency departments of Children's Medical Center of Dallas, Dallas, Tex, and Children's Hospital of the King's Daughters, Norfolk, Va, who were known or suspected to have ingested a metallic foreign body. INTERVENTION: Each patient underwent HHMD scanning and radiographic evaluation. MAIN OUTCOME MEASURES: Statistical evaluation compared HHMD scanning with radiography and experienced vs inexperienced investigator HHMD scanning to determine the accuracy of the screening tool and investigators.
RESULTS: Experienced investigators performed HHMD scans on 140 subjects; inexperienced investigators scanned all subjects. Disease was defined as a foreign body in the esophagus on radiograph. The 3 experienced investigators demonstrated sensitivity of 100% (95% confidence interval [CI], 94%-100%), specificity of 92.4% (95% CI, 84.2%-97.2%), positive predictive value (PPV) of 90.9% (95% CI, 81.3%-96.6%), and negative predictive value (NPV) of 100% (95% CI, 95.1%-100%); the proportion correct was 95.7% (95% CI, 90.8%-98.4%). The inexperienced investigators demonstrated sensitivity of 95.7% (95% CI, 88.0%-99.1%), specificity of 81.0% (95% CI, 72.1%-88.0%), PPV of 77.0% (95% CI, 66.8%-85.4%), and NPV of 96.6% (95% CI, 90.4%-99.3%); the proportion correct was 86.9% (95% CI, 80.9%-91.5%). The McNemar test demonstrated no statistically significant difference between HHMD scanning by experienced vs inexperienced investigators.
CONCLUSIONS: Handheld metal detector scanning is an accurate, inexpensive, radiation-free screening tool and should be used for evaluation of patients suspected of ingesting coins and coinlike foreign bodies.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10437760     DOI: 10.1001/archpedi.153.8.853

Source DB:  PubMed          Journal:  Arch Pediatr Adolesc Med        ISSN: 1072-4710


  8 in total

1.  Endoscopic removal of multiple duodenum foreign bodies: An unusual occurrence.

Authors:  Sameer R Islam; Ebtesam A Islam; David Hodges; Kenneth Nugent; Sreeram Parupudi
Journal:  World J Gastrointest Endosc       Date:  2010-05-16

Review 2.  Detection of coins ingested by children using a handheld metal detector: a systematic review.

Authors:  J B Lee; S Ahmad; C P Gale
Journal:  Emerg Med J       Date:  2005-12       Impact factor: 2.740

3.  Handheld Metal Detector for Metallic Foreign Body Ingestion in Pediatric Emergency.

Authors:  Hazwani Binte Hamzah; Vigil James; Suraj Manickam; Sashikumar Ganapathy
Journal:  Indian J Pediatr       Date:  2018-01-04       Impact factor: 1.967

4.  Things that go beep: experience with an ED guideline for use of a handheld metal detector in the management of ingested non-hazardous metallic foreign bodies.

Authors:  S L Ramlakhan; D P Burke; J Gilchrist
Journal:  Emerg Med J       Date:  2006-06       Impact factor: 2.740

5.  Handheld Metal Detector Screening for Metallic Foreign Body Ingestion in Children.

Authors:  Vigil James; Hazwani Binte Hamzah; Sashikumar Ganapathy
Journal:  J Vis Exp       Date:  2018-09-11       Impact factor: 1.355

6.  Acute esophageal coin ingestions: is immediate removal necessary?

Authors:  Ghazala Q Sharieff; Tonia J Brousseau; James A Bradshaw; Javaid A Shad
Journal:  Pediatr Radiol       Date:  2003-10-10

7.  The use of a hand-held metal detector for localisation of ingested metallic foreign bodies - a critical investigation.

Authors:  Johannes Schalamon; Emir Q Haxhija; Herwig Ainoedhofer; Alja Gössler; Jürgen Schleef
Journal:  Eur J Pediatr       Date:  2004-02-05       Impact factor: 3.183

8.  Management of gastric metallic foreign bodies in children.

Authors:  Andrew Au; Ran D Goldman
Journal:  Can Fam Physician       Date:  2021-07       Impact factor: 3.275

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.