Evaluation of disinfection and sterilization of reusable angioscopes with the duck hepatitis B model.

X Chaufour1, A K Deva, K Vickery, J Zou, P Kumaradeva, G H White, Y E Cossart.   

Abstract

PURPOSE: Nosocomial transmission of viral hepatitis and retrovirus infection has been reported. The expected risk is greatest for the hepatitis B virus (HBV). The duck HBV (DHBV) has similar biologic and structural characteristics to HBV and has been adopted as a suitable model for disinfectant testing.
METHODS: Angioscopic examination of the external jugular vein was performed on DHBV-infected ducks. After use, the instrument was air dried for 3 minutes. Samples were obtained by flushing the channel with 5 mL of phosphate buffered saline solution. The samples were collected immediately after drying (control), after flushing with 5 mL of water, after glutaraldehyde disinfection for 5, 10, and 20 minutes, and after ethylene oxide gas sterilization. Angioscopes were either precleaned or uncleaned before disinfection/sterilization. Residual infectivity was assessed with inoculation of samples into the peritoneal cavity of day-old ducks (n = 231).
RESULTS: DNA analysis results of liver samples showed that all 38 control ducks became infected. The frequency of DHBV infection was reduced to 93% (14 of 15) by flushing the angioscope with 5 mL of sterile water. No transmission occurred after the use of any of the properly precleaned and disinfected/sterilized angioscopes. However, after the use of the uncleaned angioscopes, the transmission rate was 90% (9 of 10) and 70% (7 of 10) after 5 and 10 minutes of contact time, respectively, in 2% glutaraldehyde. Even after the recommended 20 minutes of contact time, there was still 6% (2 of 35) transmission. After ethylene oxide sterilization, two of the recipient ducklings (2 of 35) were infected with DHBV.
CONCLUSION: There was no disease transmission after reuse of disposable angioscopes adequately cleaned before disinfection or sterilization. However, if the angioscopes are inadequately cleaned, DHBV can survive despite glutaraldehyde disinfection or ethylene oxide sterilization. This contrasts with previous in vitro and in vivo data with solid surgical instruments. It is postulated that the presence of a narrow lumen or residual protein shielding within the lumen may compromise effective inactivation of hepadnaviruses on angioscopes, with the potential risk for patient-to-patient transmission.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10436447     DOI: 10.1016/s0741-5214(99)70138-2

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  7 in total

1.  Efficacy and safety of reuse of disposable laparoscopic instruments in laparoscopic cholecystectomy: a prospective randomized study.

Authors:  T Colak; G Ersoz; T Akca; A Kanik; S Aydin
Journal:  Surg Endosc       Date:  2004-03-19       Impact factor: 4.584

Review 2.  Is peracetic acid suitable for the cleaning step of reprocessing flexible endoscopes?

Authors:  Günter Kampf; Patricia M Fliss; Heike Martiny
Journal:  World J Gastrointest Endosc       Date:  2014-09-16

Review 3.  Is hepatitis B-virucidal validation of biocides possible with the use of surrogates?

Authors:  Andreas Sauerbrei
Journal:  World J Gastroenterol       Date:  2014-01-14       Impact factor: 5.742

Review 4.  Transmission of infection by flexible gastrointestinal endoscopy and bronchoscopy.

Authors:  Julia Kovaleva; Frans T M Peters; Henny C van der Mei; John E Degener
Journal:  Clin Microbiol Rev       Date:  2013-04       Impact factor: 26.132

5.  Effective reprocessing of reusable dispensers for surface disinfection tissues - the devil is in the details.

Authors:  Günter Kampf; Stina Degenhardt; Sibylle Lackner; Christiane Ostermeyer
Journal:  GMS Hyg Infect Control       Date:  2014-03-07

6.  Does limited virucidal activity of biocides include duck hepatitis B virucidal action?

Authors:  Andreas Sauerbrei; Michael Schacke; Brigitte Glück; Uwe Bust; Holger F Rabenau; Peter Wutzler
Journal:  BMC Infect Dis       Date:  2012-10-30       Impact factor: 3.090

7.  Inactivation of feline calicivirus, a surrogate of norovirus (formerly Norwalk-like viruses), by different types of alcohol in vitro and in vivo.

Authors:  C Gehrke; J Steinmann; P Goroncy-Bermes
Journal:  J Hosp Infect       Date:  2004-01       Impact factor: 3.926

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.