We developed a new type of spacer made of hydroxyapatite (the STSS spacer) for double-door laminoplasty, and evaluated the contact of 93 STSS spacers with the split spinous processes in 20 patients with double-door laminoplasty. Contact was assessed by measuring the extent of touch of the spacer to the spinous processes, classified into four categories based on computed tomography (CT) images: excellent, complete touch on both sides of the spacer to the spinous process; good, complete touch on one side and more than half touch on the other side; fair, more than half touch on both sides; poor, half or less touch on at least one side. Excellent contact was achieved in 65 spacers (69.9%); good, in 13 (14.0%); fair, in 11 (11. 8%); and poor, in 4 (4.3%). The percentages of excellent or good categories were 75.0% at the C3 level, 73.7% at the C4 level, 78.9% at the C5 level, 90.0% at the C6 level, and 100% at the C7 level. The contact rate of the STSS spacer with the spinous process was better than that achieved with other spacers, probably because the characteristic shape of the STSS spacer was compatible with the widened space between the bilateral spinous processes; i.e., it is trapezoidal on both the axial and the frontal sections. However, the appropriate size of the spacer must be selected in accordance with the size of the spinous process to obtain higher percentages of excellent or good contact.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10436273     DOI: 10.1007/s007760050102

Source DB:  PubMed          Journal:  J Orthop Sci        ISSN: 0949-2658            Impact factor:   1.601


  8 in total

1.  Comparison of enlargement of the spinal canal after cervical laminoplasty: open-door type and double-door type.

Authors:  Shigeru Hirabayashi; Hironobu Yamada; Takao Motosuneya; Yoshinobu Watanabe; Makoto Miura; Hiroya Sakai; Takashi Matsushita
Journal:  Eur Spine J       Date:  2010-03-23       Impact factor: 3.134

2.  Midline-Splitting Open Door Laminoplasty Using Hydroxyapatite Spacers : Comparison between Two Different Shaped Spacers.

Authors:  Jin Hoon Park; Sang Ryong Jeon
Journal:  J Korean Neurosurg Soc       Date:  2012-07-31

3.  Midline Splitting Cervical Laminoplasty Using Allogeneic Bone Spacers: Comparison of Fusion Rates between Cervical Spondylotic Myelopathy and Ossification of Posterior Longitudinal Ligament.

Authors:  Jae Jon Sheen; Sang Ryong Jeon
Journal:  Korean J Neurotrauma       Date:  2014-10-31

4.  Novel Hybrid Hydroxyapatite Spacers Ensure Sufficient Bone Bonding in Cervical Laminoplasty.

Authors:  Nobuhiro Tanaka; Kazuyoshi Nakanishi; Naosuke Kamei; Toshio Nakamae; Shinji Kotaka; Yoshinori Fujimoto; Mitsuo Ochi; Nobuo Adachi
Journal:  Asian Spine J       Date:  2018-10-16

5.  Recent Surgical Methods of Double-door Laminoplasty of the Cervical Spine (Kurokawa's Method).

Authors:  Shigeru Hirabayashi
Journal:  Spine Surg Relat Res       Date:  2018-02-28

6.  The Effect of Mesenchymal Stem Cell-Enriched Scaffolds on MMP-8 and TGF-β Levels of Vertebrae Postlaminoplasty in Rabbit Model.

Authors:  Ahmad Jabir Rahyussalim; Roni Eka Sahputra; Menkher Manjas; Yudan Whulanza; Tri Kurniawati; Dina Aprilya; Muhammad Luqman Labib Zufar
Journal:  Stem Cells Cloning       Date:  2021-07-12

Review 7.  Two types of laminoplasty for cervical spondylotic myelopathy at multiple levels.

Authors:  Shigeru Hirabayashi; Takashi Matsushita
Journal:  ISRN Orthop       Date:  2011-09-07

8.  Improved bone bonding of hydroxyapatite spacers with a high porosity in a quantitative computed tomography-image pixel analysis: A prospective 1-year comparative study of the consecutive cohort undergoing double-door cervical laminoplasty.

Authors:  Yoshiki Takeoka; Takashi Yurube; Koichiro Maeno; Yutaro Kanda; Ryu Tsujimoto; Kunihiko Miyazaki; Yuji Kakiuchi; Shingo Miyazaki; Zhongying Zhang; Toru Takada; Kotaro Nishida; Minoru Doita; Ryosuke Kuroda; Kenichiro Kakutani
Journal:  JOR Spine       Date:  2020-02-03
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.