Literature DB >> 10435532

Dependence of scatter on atomic number for x rays from tungsten and molybdenum anodes in the mammographic energy range.

R J Aus1, L A DeWerd, D W Pearson, J A Micka, K H Ng.   

Abstract

A study was done to determine the relative amounts of scatter for the following materials with atomic numbers ranging from Z=6 to Z=82: C, Al, Ti, Fe, Cu, Zn, Zr, Y, Mo, Ta, and Pb. Measurements were performed for each material on two constant potential x-ray units--one fitted with a molybdenum (Mo) anode-Mo filter and the other with a tungsten (W) anode-aluminum (Al) filter (medium filtration) at 30 kVp. Theoretical calculations were also performed for each anode to explain the scatter behavior and to aid in predicting the behavior for materials where measurements were not made. There was good agreement between the theoretical calculations and the experimental data.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10435532     DOI: 10.1118/1.598625

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071