Calculated vs measured pharyngeal mucosal pressures with the laryngeal mask airway during cuff inflation: assessment of four locations.

C Keller1, J Brimacombe, A Benzer.   

Abstract

We have compared calculated with measured pharyngeal mucosal pressures at four different locations on the surface of the laryngeal mask airway (LMA) during cuff inflation in 10 anaesthetized, paralysed adult patients. Microchip sensors were attached to a size 5 LMA at the following locations: the anterior and lateral side, tip and backplate. Pressures were recorded during inflation of the cuff from 0 to 40 ml in 5-ml increments. Calculated pressures were determined by subtracting in vivo from in vitro intracuff pressures. Calculated pressures were greater than measured pressures at cuff volumes of 5 ml or greater at all locations (P < or = 0.003). The greatest mean calculated the measured pressures were 118 and 14 cm H2O, respectively. We conclude that measured mucosal pressures at the four locations tested were less than calculated pressures and less than capillary perfusion pressure.

Entities:  

Mesh:

Year:  1999        PMID: 10434824     DOI: 10.1093/bja/82.3.399

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  5 in total

1.  [Continuous cuff pressure measurement during laryngeal mask anesthesia : An obligatory measure to avoid postoperative complications].

Authors:  M Hensel; T Güldenpfennig; A Schmidt; M Krumm
Journal:  Anaesthesist       Date:  2016-04-12       Impact factor: 1.041

2.  [Laryngeal tube versus laryngeal mask airway in anaesthetised non-paralysed patientsA comparison of handling and postoperative morbidity].

Authors:  M Wrobel; U Grundmann; W Wilhelm; S Wagner; R Larsen
Journal:  Anaesthesist       Date:  2004-08       Impact factor: 1.041

Review 3.  Complications Associated with the Use of Supraglottic Airway Devices in Perioperative Medicine.

Authors:  Pavel Michalek; William Donaldson; Eliska Vobrubova; Marek Hakl
Journal:  Biomed Res Int       Date:  2015-12-13       Impact factor: 3.411

4.  Correlation and variation of cuff inflating volumes and pressures in different adult models of laryngeal mask: a prospective randomized trial.

Authors:  Narut Ruananukun; Jittiya Watcharotayangul; Suchaya Jeeranukosol; Rojnarin Komonhirun
Journal:  BMC Anesthesiol       Date:  2020-05-07       Impact factor: 2.217

5.  Application of Minimum Effective Cuff Inflating Volume for Laryngeal Mask Airway and its Impact on Postoperative Pharyngeal Complications.

Authors:  Bing-Bing Li; Jie Yan; Hong-Gang Zhou; Jing Hao; Ai-Jia Liu; Zheng-Liang Ma
Journal:  Chin Med J (Engl)       Date:  2015-10-05       Impact factor: 2.628

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.