Extravasation detection accessory: clinical evaluation in 500 patients.

B A Birnbaum1, R C Nelson, J L Chezmar, S N Glick.   

Abstract

PURPOSE: To assess the ability of an extravasation detection accessory (EDA) to detect clinically important (> or = 10 mL) extravascular injection of iodinated contrast material delivered with an automated power injector.
MATERIALS AND METHODS: Five hundred patients referred for contrast material-enhanced body computed tomography (CT) participated in a prospective, multi-institutional, observational study in which the EDA was used to identify and interrupt any injection associated with clinically important extravasation. The presence or absence of extravasation was definitively established with helical CT at the injection site (injection rate, from 1.0 to 5.0 mL/sec; mean, 2.9 mL/sec; median, 3.0 mL/sec).
RESULTS: There were four true-positive (extravasation volumes, 13-18 mL), 484 true-negative, 12 false-positive, and no false-negative cases. The prevalence of overall and clinically important (> or = 10 mL) extravasation was 3.6% (18 of 500 cases) and 0.8% (four of 500 cases), respectively. The EDA had a sensitivity of 100% (four of four cases; 95% CI: 51%, 100%) and a specificity of 98% (484 of 496 cases; 95% CI: 96%, 99%) in the detection of clinically important extravasation.
CONCLUSION: The EDA is easy to use, safe, and accurate in the monitoring of intravenous injections for extravasation, which could prove especially useful in high-flow-rate CT applications.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10429700     DOI: 10.1148/radiology.212.2.r99au14431

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  4 in total

1.  Contrast media extravasations in patients undergoing computerized tomography scanning: a systematic review and meta-analysis of risk factors and interventions.

Authors:  Sandrine Ding; Nicole Richli Meystre; Cosmin Campeanu; Giuseppe Gullo
Journal:  JBI Database System Rev Implement Rep       Date:  2018-01

2.  Intravenous contrast medium extravasation: systematic review and updated ESUR Contrast Media Safety Committee Guidelines.

Authors:  Giles Roditi; Nadir Khan; Aart J van der Molen; Marie-France Bellin; Michele Bertolotto; Torkel Brismar; Jean-Michel Correas; Ilona A Dekkers; Remy W F Geenen; Gertraud Heinz-Peer; Andreas H Mahnken; Carlo C Quattrocchi; Alexander Radbruch; Peter Reimer; Laura Romanini; Fulvio Stacul; Henrik S Thomsen; Olivier Clément
Journal:  Eur Radiol       Date:  2022-02-17       Impact factor: 7.034

3.  Acute compartment syndrome of hand resulting from radiographic contrast iohexol extravasation.

Authors:  Kolar Vishwanath Vinod; Rampelli Shravan; Radhakrishnan Shrivarthan; Pedapati Radhakrishna; Tarun Kumar Dutta
Journal:  J Pharmacol Pharmacother       Date:  2016 Jan-Mar

4.  Biceps Brachii Compartment Contrast Media Extravasation with Surgical Correlation.

Authors:  Felix S Chew; Carol A Boles; Christopher T Mattern
Journal:  Radiol Case Rep       Date:  2015-11-06
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.