Relative dose uniformity assessment in interstitial implants.

V Y Wong1, T W Leung, C M Wong.   

Abstract

PURPOSE: Two new indices, the peak index (PI) and the new geometry index (NGI), that quantify implant dose uniformity and quality are presented. Their advantages include independence to absolute treatment dose and high sensitivity compared with other adopted dose-uniformity measures. The applicability of these indices were evaluated through computer simulations and several clinically executed implant cases. Target coverage is assumed to be properly observed and will not be discussed herein. METHODS AND MATERIALS: The natural volume-dose histogram serves as the basis of our investigation. The PI and NGI definitions are based on parameters derived from the histogram. Two computer-simulated implants and 12 clinically executed implants, using high-dose rate remote afterloading techniques, are studied. Various indices that quantify the dose uniformity of the implant, namely the quality index (QI), geometry index, as well as the PI and NGI, are computed, and the results are compared.
RESULTS: The PI demonstrated significantly increased sensitivity (up to 5 times) to dose-uniformity evaluation, compared with the QI. The deduced parameter NGI may thus offer a better measure of implant qualities, allowing a more meaningful assessment and correlation between implant qualities to the treatment results. The PI system also offers a guideline to the design of optimal implant geometry.
CONCLUSION: The PI overcomes some of the shortcomings of the QI in that it provides more information about the peaking of the natural dose-volume histogram of a particular implant. The PI and NGI may offer better, more sensitive means to assess implant dose uniformity, independent of prescription dose, than other measures.

Mesh:

Year:  1999        PMID: 10421553     DOI: 10.1016/s0360-3016(99)00129-7

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  2 in total

1.  Quality assessment of interstitial implants in high- dose- rate brachytherapy after lumpectomy in patients of early stage breast cancer.

Authors:  Samrat Dutta; Suresh C Sharma; Rajinder Singh; Rakesh Kapoor; Arun S Oinam; Anoop Bhardwaj; Pradeep Goswami
Journal:  Indian J Surg Oncol       Date:  2011-03-24

2.  Assessment of dose homogeneity in conformal interstitial breast brachytherapy with special respect to ICRU recommendations.

Authors:  Tibor Major; Georgina Fröhlich; Csaba Polgar
Journal:  J Contemp Brachytherapy       Date:  2011-09-30
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.