Literature DB >> 10420368

[Measuring choroid blood flow with a new confocal laser Doppler device].

M H Geiser1, C E Riva, U Diermann.   

Abstract

UNLABELLED: A new instrument for the measurement of choroidal blood flow in the fovea is presented. It is based on the laser Doppler method and a confocal optical system with an indirect detection of the Doppler shifted light.
METHOD: The intensity of the laser beam (785 nm) at the cornea is 90 microW. Measurements were obtained from a normal population of 21 subjects under resting conditions without dilating the pupil.
RESULTS: The reproducibility of the choroidal blood flow, based on 5 measurements of 10 s each in 5 randomly selected subjects, is 9%. The minimum detectable change for a statistical significance of p < or = 0.05, based on a population of 21 subjects and 10 s measurements, is 9%.
CONCLUSION: This new compact instrument appears to be suitable for the investigation of the physiology and pharmacology of choroidal blood flow and the effect of age-related macular degeneration.

Mesh:

Year:  1999        PMID: 10420368     DOI: 10.1055/s-2008-1034795

Source DB:  PubMed          Journal:  Klin Monbl Augenheilkd        ISSN: 0023-2165            Impact factor:   0.700