Evaluation and implementation of the gel test for indirect antiglobulin testing in a community hospital laboratory.

J C Cate1, N Reilly.   

Abstract

BACKGROUND: The gel test, developed by Lapierre in 1984, was designed to standardize antiglobulin testing while improving sensitivity and specificity of the method. PRINCIPLE: Anti-human serum immunoglobulin G (IgG) mixed with Sephadex G100 (gel phase) in a microtube traps red cell-IgG agglutination complexes during migration through the gel in a centrifugation step. Agglutination complexes are visibly detectable at various levels in the microtube as an inverse function of antibody coated on red cells. Unsensitized red cells form a cell pellet at the base of the microtube.
OBJECTIVE: To determine if indirect anti-human globulin testing could be standardized and simplified by replacing the tube test with the gel test without compromising quality or increasing costs.
SETTING: A medium-sized community hospital.
RESULTS: In a blinded retrospective study, we used patient sera (n = 40), which included 10 positive specimens containing 18 known antibodies. Sixteen antibodies were detected and identified with the tube method (1 anti-D and 1 anti-C not detected). By the gel test, 18 antibodies were detected and identified. All negative samples showed 100% concordance. Favorable results were obtained in a nonblinded prospective correlation study (n = 121). Our technologists found the gel test easier to read and more reproducible and reliable than the tube method; they also found increased sensitivity for detecting weakly reacting antibodies. We successfully introduced the gel test into our laboratory as the standard method for indirect antiglobulin testing. Following implementation, improved personnel management was achieved.
CONCLUSIONS: The gel test is a reliable and advantageous method and is appropriate for routine use for detection and identification of alloantibodies in a community hospital transfusion service laboratory.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10420225     DOI: 10.5858/1999-123-0693-EAIOTG

Source DB:  PubMed          Journal:  Arch Pathol Lab Med        ISSN: 0003-9985            Impact factor:   5.534


  6 in total

1.  Comparative Study of Blood Cross Matching Using Conventional Tube and Gel Method.

Authors:  D Swarup; P S Dhot; J Kotwal; A K Verma
Journal:  Med J Armed Forces India       Date:  2011-07-21

2.  Antibody elutions in Thai patients with a positive direct antiglobulin test.

Authors:  Oytip Nathalang; Pramote Sriwanitchrak; Jintana Tubrod; Pawinee Kupatawintu
Journal:  Blood Transfus       Date:  2010-09-13       Impact factor: 3.443

3.  Good agreement of conventional and gel-based direct agglutination test in immune-mediated haemolytic anaemia.

Authors:  Christine J Piek; Erik Teske; Martin W van Leeuwen; Michael J Day
Journal:  Acta Vet Scand       Date:  2012-02-08       Impact factor: 1.695

4.  Red cell phenotyping of blood from donors at the National blood center of Malaysia.

Authors:  Rozi H Musa; Suhair A Ahmed; Hasna Hashim; Yasmin Ayob; Nor H Asidin; Poh Y Choo; Fawwaz S Al-Joudi
Journal:  Asian J Transfus Sci       Date:  2012-01

5.  Comparison of micro column technology with conventional tube methods for antibody detection.

Authors:  Sachin Garg; Nishant Saini; Ravneet Kaur Bedi; Sabita Basu
Journal:  J Lab Physicians       Date:  2017 Apr-Jun

6.  Convolutional neural network-based automatic classification for incomplete antibody reaction intensity in solid phase anti-human globulin test image.

Authors:  KeQing Wu; ShengBao Duan; YuJue Wang; HongMei Wang; Xin Gao
Journal:  Med Biol Eng Comput       Date:  2022-03-07       Impact factor: 3.079

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.