| Literature DB >> 10410953 |
M Mizugaki1, T Hishinuma, N Suzuki.
Abstract
A liquid chromatographic-tandem mass spectrometric (LC-MS-MS) method was developed for the quantitation of urinary leukotriene E4 (LTE4). LTE4 and its internal standard were extracted by solid-phase extraction and analysed using LC-MS-MS in the selected reaction monitoring (SRM) mode. A good linear response over the range of 10 pg to 10 ng was demonstrated. The accuracy of added LTE4 ranged from 97.0% to 108.0% with a mean and SD of 100.6+/-2.4%. We detected LTE4 (63.1+/-18.7 pg/mg creatinine, n=10) in healthy human urine. This method can be used to determine LTE4 in biological samples.Entities:
Mesh:
Substances:
Year: 1999 PMID: 10410953 DOI: 10.1016/s0378-4347(99)00174-7