Severe accidental hypothermia: rewarming strategy using a veno-venous bypass system and a convective air warmer.

A Bräuer1, H Wrigge, J Kersten, J Rathgeber, W Weyland, H Burchardi.   

Abstract

OBJECTIVE: To study a rewarming strategy for patients with severe accidental hypothermia using a simple veno-venous bypass in combination with a convective air warmer.
SETTING: Eighteen beds in a university hospital intensive care unit. PATIENTS: Four adults admitted with a core temperature less than 30 degrees C. Hypothermia was caused by alcoholic intoxication in three patients and by drug overdose in one patient.
MEASUREMENTS AND MAIN RESULTS: All patients were rewarmed by a venovenous bypass and in three cases a convective air warmer was also used. At a bypass flow rate of 100-300 ml/min the mean increase in core temperature was 1.15 degrees C/h (Range: 1.1-1.2 degrees C/h). One patient died 2 days after rewarming as a consequence of a reactivated pancreatitis. The other three patients survived without neurological sequelae.
CONCLUSION: This rewarming technique seems safe and effective and allowed the controlled rewarming of our patients who suffered from severe accidental hypothermia

Entities:  

Mesh:

Year:  1999        PMID: 10401949     DOI: 10.1007/s001340050891

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  4 in total

1.  [Rewarming methods for severely injured hypothermic patients].

Authors:  P Kobbe; P Lichte; M Wellmann; F Hildebrand; D Nast-Kolb; C Waydhas; R Oberbeck
Journal:  Unfallchirurg       Date:  2009-12       Impact factor: 1.000

2.  Active intravascular rewarming for hypothermia associated with traumatic injury: early experience with a new technique.

Authors:  Edward E Taylor; James P Carroll; Matthew A Lovitt; Laura B Petrey; Paul E Gray; Cyndi J Mastropieri; Michael L Foreman
Journal:  Proc (Bayl Univ Med Cent)       Date:  2008-04

Review 3.  [Management of accidental hypothermia].

Authors:  M Hohlrieder; M Kaufmann; M Moritz; V Wenzel
Journal:  Anaesthesist       Date:  2007-08       Impact factor: 1.041

4.  Iceman Survived due to Cooling Device.

Authors:  M Roser; F Martens; C Storm
Journal:  ISRN Cardiol       Date:  2011-05-09
  4 in total