Literature DB >> 10395270

Long-term patency rates, complications and cost-effectiveness of polytetrafluoroethylene (PTFE) grafts for hemodialysis access: a prospective study that compares Impra versus Gore-tex grafts.

S N Hurlbert1, M A Mattos, J P Henretta, D E Ramsey, L D Barkmeier, K J Hodgson, D S Summer.   

Abstract

Manufacturers of polytetraflouroethylene (PTFE) grafts used for chronic hemodialysis access describe specific advantages for their respective grafts, which presumably result in greater graft patency rates, reduced complications and decreased overall costs. There are few data available in the literature to support or contradict these alleged benefits. Therefore, this prospective study was undertaken to evaluate and compare patency rates, complications and costs between two of the leading brands of PTFE that are currently being marketed for use as hemodialysis access grafts. Totals of 190 primary PTFE grafts (100 Gore-tex (W. L. Gore and Associates, Flagstaff, AZ) and 90 Impra (C. R. Bard Inc., Tempe, AZ)) were implanted in 168 consecutive patients with end-stage renal disease. A policy of non-interventions was employed for patent grafts, as no attempt was made to assist primary patency. Grafts that occluded during follow-up underwent secondary revision to maintain patency. There was no difference in primary and secondary patency by life-table analysis between Gore-tex and Impra grafts at 2 years (P > 0.53 and P > 0.13, respectively). There was also no significant difference between Gore-tex and Impra in the number of days before the first thrombectomy or in the number of thrombectomies or revisions per graft (P > O.50). Likewise, the incidence of complications was similar between the two grafts. The cost of graft implantation and maintenance of patency was not significantly different between Gore-tex and Impra grafts. It is concluded that either graft can be used for hemodialysis access with similar expected outcomes for at least 2 years following implantation.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 10395270     DOI: 10.1016/s0967-2109(98)00062-3

Source DB:  PubMed          Journal:  Cardiovasc Surg        ISSN: 0967-2109


  4 in total

1.  Patency and life-spans of failing hemodialysis grafts in patients undergoing repeated percutaneous de-clotting.

Authors:  A V Mansilla; B D Toombs; W K Vaughn; J I Zeledon
Journal:  Tex Heart Inst J       Date:  2001

Review 2.  [Vascular grafts as access for hemodialysis].

Authors:  J Zanow; U Settmacher
Journal:  Chirurg       Date:  2012-09       Impact factor: 0.955

3.  Outcomes of two different polytetrafluoroethylene graft sizes in patients undergoing maintenance hemodialysis.

Authors:  R Afshar; S Sanavi; S Afshin-Majd; A Davati
Journal:  Indian J Nephrol       Date:  2009-10

4.  Patency rate and complications of polytetrafluoroethylene grafts compared with polyurethane grafts for hemodialysis access.

Authors:  Hassan Ravari; Gholam Hossein Kazemzade; Mohammad Hadi Saied Modaghegh; Patricia Khashayar
Journal:  Ups J Med Sci       Date:  2010-03-11       Impact factor: 2.384

  4 in total