A simple radiographic measurement method for polyethylene wear in total knee arthroplasty.

Y Fukuoka1, A Hoshino, A Ishida.   

Abstract

This study describes a new method for evaluating polyethylene wear in total knee arthroplasty. Since the amount of wear is dependent on a number of variables such as the weight and activity of the patient, it should be estimated based on in vivo measurements. We used a computer vision technique called three-dimensional/two-dimensional (3-D/2-D) matching to perform in vivo assessment using a single-plane radiograph. Using the 3-D/2-D matching algorithm we estimated the 3-D position and orientation of each knee implant and then measured the femorotibial distance, which is defined as the shortest perpendicular distance from the tibial tray to the femoral component. The accuracy of the proposed 3-D/2-D matching method was determined by in vitro investigations. The worst errors in in-plane/out-of-plane translations and rotations were 0.20/1.95 mm and 0.17/0.29 degrees, respectively. The root-mean-square error in femorotibial distance measurements using real polyethylene inserts was 0.04 mm. Results of in vivo femorotibial distance measurements are also described.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10391593     DOI: 10.1109/86.769413

Source DB:  PubMed          Journal:  IEEE Trans Rehabil Eng        ISSN: 1063-6528


  4 in total

1.  A 3D kinematic estimation of knee prosthesis using X-ray projection images: clinical assessment of the improved algorithm for fluoroscopy images.

Authors:  Shunji Hirokawa; M Abrar Hossain; Yuichi Kihara; Shogo Ariyoshi
Journal:  Med Biol Eng Comput       Date:  2008-09-30       Impact factor: 2.602

2.  Predicting 3D pose in partially overlapped X-ray images of knee prostheses using model-based Roentgen stereophotogrammetric analysis (RSA).

Authors:  Chi-Pin Hsu; Shang-Chih Lin; Kao-Shang Shih; Chang-Hung Huang; Chian-Her Lee
Journal:  Med Biol Eng Comput       Date:  2014-10-08       Impact factor: 2.602

3.  Pose estimation of artificial hip joint using a single radiographic projection.

Authors:  H Inaoka; A Ishidal; Y Fukuoka; K Suzuki; M Matsubara
Journal:  Med Biol Eng Comput       Date:  2003-01       Impact factor: 2.602

4.  Weight of polyethylene wear particles is similar in TKAs with oxidized zirconium and cobalt-chrome prostheses.

Authors:  Young-Hoo Kim; Jun-Shik Kim; Wansoo Huh; Kwang-Hoon Lee
Journal:  Clin Orthop Relat Res       Date:  2009-12-01       Impact factor: 4.176