Experience with a programmable valve shunt system.

N Yamashita1, K Kamiya, K Yamada.   

Abstract

OBJECT: The goal of this study was to clarify the efficacy of the Codman Hakim programmable valve. Clinical data obtained in 179 patients with hydrocephalus or other intracranial fluid-accumulating diseases were analyzed.
METHODS: Shunt placement operations were effective in 168 patients, approximately one half (50.6%) of whose devices required reprogramming of opening pressure postoperatively. This was a significantly larger number of shunts than the authors had thought would need reprogramming. Extremely narrowed ventricles observed on computerized tomography scans, as well as clinical symptoms related to inadequate or excessive cerebrospinal fluid drainage, improved in patients after shunt reprogramming. Shunt reprogramming frequently was necessary in patients with posthemorrhagic acute hydrocephalus; the programmable valve proved particularly beneficial for such patients. Subdural effusion and arachnoid cyst also proved to be good indications for use of the valve. Twelve patients (7%) suffered complications postoperatively. The most common complication was valve obstruction, which occurred in five patients, most of whom had brain tumors.
CONCLUSIONS: The programmable valve was beneficial for the treatment of hydrocephalus and other intracranial fluid-accumulating diseases. It is important to be careful in selecting patients for treatment with the programmable valve, because complications involving the valve seem more likely in brain tumor cases. The valve proved to be poorly resistant to magnetic fields; therefore, it is essential to confirm opening pressure after every magnetic resonance imaging examination. The authors recommend that an identification system for patients be developed so that medical personnel will be aware of the presence of the valve and the previous setting of opening pressure.

Entities:  

Mesh:

Year:  1999        PMID: 10389876     DOI: 10.3171/jns.1999.91.1.0026

Source DB:  PubMed          Journal:  J Neurosurg        ISSN: 0022-3085            Impact factor:   5.115


  8 in total

1.  Is there a correlation between operative results and change in ventricular volume after shunt placement? A study of 60 cases of idiopathic normal-pressure hydrocephalus.

Authors:  U Meier; S Paris; A Gräwe; D Stockheim; A Hajdukova; S Mutze
Journal:  Neuroradiology       Date:  2003-05-16       Impact factor: 2.804

2.  The Strata programmable valve for shunt-dependent hydrocephalus: the pediatric experience at a single institution.

Authors:  Edward S Ahn; Markus Bookland; Benjamin S Carson; Jon D Weingart; George I Jallo
Journal:  Childs Nerv Syst       Date:  2006-10-07       Impact factor: 1.475

Review 3.  Shunt Devices for Neurointensivists: Complications and Management.

Authors:  G Smith; J Pace; A Scoco; G Singh; K Kandregula; S Manjila; C Ramos-Estebanez
Journal:  Neurocrit Care       Date:  2017-10       Impact factor: 3.210

Review 4.  Is there an ideal shunt? A panoramic view of 110 years in CSF diversions and shunt systems used for the treatment of hydrocephalus: from historical events to current trends.

Authors:  Nigel Peter Symss; Shizuo Oi
Journal:  Childs Nerv Syst       Date:  2014-12-30       Impact factor: 1.475

Review 5.  Cerebrospinal fluid hydrocephalus shunting: cisterna magna, ventricular frontal, ventricular occipital.

Authors:  Seifollah Gholampour; Jay Patel; Bakhtiar Yamini; David Frim
Journal:  Neurosurg Rev       Date:  2022-05-05       Impact factor: 2.800

6.  The treatment of large supratentorial arachnoid cysts in infants with cyst-peritoneal shunting and Hakim programmable valve.

Authors:  Antonino Germanò; Gerardo Caruso; Mariella Caffo; Sergio Baldari; Amedeo Calisto; Francesco Meli; Francesco Tomasello
Journal:  Childs Nerv Syst       Date:  2003-02-13       Impact factor: 1.475

7.  Management of neonatal hydrocephalus: feasibility of use and safety of two programmable (Sophy and Polaris) valves.

Authors:  Juan F Martínez-Lage; María-José Almagro; Isabel Sanchez Del Rincón; Miguel A Pérez-Espejo; Claudio Piqueras; Raúl Alfaro; Javier Ros de San Pedro
Journal:  Childs Nerv Syst       Date:  2007-10-09       Impact factor: 1.475

8.  Shunt Overdrainage Caused by Displacement of the Pressure Control Cam after Pressure Adjustment.

Authors:  Su-Ho Kim; Yoon-Soo Lee; Min-Seok Lee; Sang-Jun Suh; Jeong-Ho Lee; Dong-Gee Kang
Journal:  Korean J Neurotrauma       Date:  2016-10-31