Literature DB >> 10379509

Technical management of a pregnant patient undergoing radiation therapy to the head and neck.

M B Podgorsak1, R J Meiler, H Kowal, S P Kishel, J B Orner.   

Abstract

The fetal dose in a pregnant patient undergoing radiation therapy to the head and neck region was investigated. Implicit in this study was the design and evaluation of a shield used to minimize the fetal dose. To evaluate the fetal dose, a phantom was irradiated with the fields designed for this patient's therapy. The peripheral dose was measured for each field individually, both without and with a custom shield designed to be placed about the patient's abdominal and pelvic regions. The total dose at the location of the fetus over the course of this patient's radiation therapy was then estimated from peripheral dose rate measurements made at several points within the simulated uterus. With no shielding, the total dose within the uterus of the patient would have ranged from 13.3 cGy at the cervix to 28 cGy at the fundus. With the shield applied, the uterine dose was significantly less: 3.3 cGy at the cervix to 8.6 cGy at the fundus. In fact, at every measurement point, the peripheral dose with the shield in place was 30% to 50% of the dose without the shield. Some data suggest that the rate of significant abnormalities induced by irradiation in utero increases with increasing dose within the range of total peripheral doses incurred during most radiation treatment courses. It is therefore prudent to make reasonable attempts at minimizing the dose to the lower abdominal and pelvic regions of any pregnant patient. The shield designed in this work accomplished this goal for this patient and is flexible enough to be used in the treatment of almost all tumor volumes.

Entities:  

Mesh:

Year:  1999        PMID: 10379509     DOI: 10.1016/s0958-3947(99)00010-2

Source DB:  PubMed          Journal:  Med Dosim        ISSN: 1873-4022            Impact factor:   1.482


  3 in total

1.  SAF values for internal photon emitters calculated for the RPI-P pregnant-female models using Monte Carlo methods.

Authors:  C Y Shi; X George Xu; Michael G Stabin
Journal:  Med Phys       Date:  2008-07       Impact factor: 4.071

2.  Specific absorbed fractions for internal electron emitters derived for a set of anatomically realistic reference pregnant female models.

Authors:  Bingqi Guo; X George Xu; Chengyu Shi
Journal:  Radiat Prot Dosimetry       Date:  2009-09-18       Impact factor: 0.972

3.  Revisiting fetal dose during radiation therapy: evaluating treatment techniques and a custom shield.

Authors:  Amir M Owrangi; Donald A Roberts; Elizabeth L Covington; James A Hayman; Kathryn M Masi; Choonik Lee; Jean M Moran; Joann I Prisciandaro
Journal:  J Appl Clin Med Phys       Date:  2016-09-08       Impact factor: 2.102