Literature DB >> 10377320

Initial experience in the extraction of chronically implanted pacemaker leads using the Excimer laser sheath.

T Levy1, S Walker, V Paul.   

Abstract

OBJECTIVE: To assess the safety and efficiency of the Excimer laser sheath in extracting chronically implanted pacemaker leads. PATIENTS: Eight patients were studied (one female, mean age 62 years, range 34 to 77) with 17 pacemaker leads (five atrial, 10 ventricular, two implantable defibrillator). The mean implantation time was 65 months (range 23 to 188). The indications for lead extraction were chronic infection (7), superior vena cava obstruction (4), lead malfunction (4), and pain (2).
METHODS: A prospective analysis of the use of the Excimer laser sheath in extracting chronically implanted pacemaker leads. Laser sheath extraction was undertaken if conventional extraction techniques with simple traction or traction with a locking stylet had failed. If laser sheath extraction was unsuccessful, basket retrieval of the lead from the groin was performed.
RESULTS: Complete lead removal was achieved in 16 leads (94%). In one case the electrode tip was left behind without complication. Extraction was achieved with the laser sheath alone in 16 leads. Basket retrieval was required in one case after laser failure. There were no complications.
CONCLUSIONS: The Excimer laser sheath appears to be an effective and safe technique for extracting chronically implanted pacemaker leads. It can be used in combination with the currently available techniques for successful lead extraction.

Entities:  

Mesh:

Year:  1999        PMID: 10377320      PMCID: PMC1729106          DOI: 10.1136/hrt.82.1.101

Source DB:  PubMed          Journal:  Heart        ISSN: 1355-6037            Impact factor:   5.994


  2 in total

1.  Pacemaker wires. Difference in performance between AMBER and conventional chest radiography.

Authors:  S G Fransson
Journal:  Acta Radiol       Date:  1993-07       Impact factor: 1.990

2.  First European experience using excimer laser for the extraction of permanent pacemaker leads.

Authors:  C Kennergren
Journal:  Pacing Clin Electrophysiol       Date:  1998-01       Impact factor: 1.976

  2 in total
  2 in total

1.  The use of radiofrequency catheter ablation to extract a chronic permanent pacemaker lead after failed laser extraction.

Authors:  Deepak Roshan Talreja; Samuel Asirvatham; David L Hayes
Journal:  J Interv Card Electrophysiol       Date:  2002-06       Impact factor: 1.900

2.  Reuse of occluded veins during permanent pacemaker lead extraction: a new indication for femoral lead extraction.

Authors:  Andrew D Staniforth; Richard J Schilling
Journal:  Indian Pacing Electrophysiol J       Date:  2002-10-01
  2 in total