Literature DB >> 10376915

Verification of linear lesions using a noncontact multielectrode array catheter versus conventional contact mapping techniques.

B Schumacher1, W Jung, T Lewalter, C Wolpert, B Lüderitz.   

Abstract

INTRODUCTION: Creation of linear lesions is an established ablation goal. Verification of complete conduction block at the ablation line is required to determine ablation success. Conventional mapping techniques are sequential endocardial activation mapping and documentation of double potentials. Recently, a noncontact multielectrode array catheter was developed that allows instantaneous three-dimensional mapping by simultaneous reconstruction of > 3,000 electrograms. In this study, we prospectively compared the accuracy of noncontact mapping to identify discontinuities in linear lesions and to verify a conduction block with that of conventional mapping techniques. METHODS AND
RESULTS: In 12 patients with atrial flutter, radiofrequency pulses were applied between the tricuspid annulus and either the inferior vena cava or the eustachian ridge. Following each application, pulse propagation at the ablation line was determined during pacing by conventional mapping techniques. The findings were compared to high-density isopotential mapping using the noncontact multielectrode array catheter. It was found that noncontact mapping reliably distinguished conduction delays from a conduction block as defined by contact mapping. In addition, noncontact mapping instantaneously identified the area where a discontinuity in the line of block was present. In these patients, complete conduction block was achieved by radiofrequency pulses guided by the noncontact mapping system.
CONCLUSION: Noncontact mapping is highly accurate in distinguishing conduction delays from a complete conduction block. By providing an instantaneous high-density propagation vector at all sites along the ablation line, three-dimensional isopotential mapping is helpful in localizing discontinuities of linear lesions and, thus, may facilitate the creation of a complete conduction block.

Entities:  

Mesh:

Year:  1999        PMID: 10376915     DOI: 10.1111/j.1540-8167.1999.tb00258.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  5 in total

1.  Benefits of new catheter mapping technology: fact or fiction?

Authors:  S Saksena
Journal:  J Interv Card Electrophysiol       Date:  2000-06       Impact factor: 1.900

2.  Noncontact mapping for radiofrequency ablation of complex cardiac arrhythmias.

Authors:  J Sra; A Bhatia; D Krum; M Akhtar
Journal:  J Interv Card Electrophysiol       Date:  2001-09       Impact factor: 1.900

3.  Identification of extremely slow conduction in the cavotricuspid isthmus during common atrial flutter ablation.

Authors:  Jian Chen; Christian de Chillou; Per Ivar Hoff; Ole Rossvoll; Marius Andronache; Nicolas Sadoul; Isabelle Magnin-Poull; Knut Ståle Erga; Etienne Aliot; Ole-Jørgen Ohm
Journal:  J Interv Card Electrophysiol       Date:  2002-08       Impact factor: 1.900

4.  Organized atrial tachycardias after atrial fibrillation ablation.

Authors:  Sergio Castrejón-Castrejón; Marta Ortega; Armando Pérez-Silva; David Doiny; Alejandro Estrada; David Filgueiras; José L López-Sendón; José L Merino
Journal:  Cardiol Res Pract       Date:  2011-09-19       Impact factor: 1.866

5.  Utility of virtual unipolar electrogram morphologies to detect transverse conduction block and turnaround points of typical atrial flutter.

Authors:  Mitsuru Takami; Akihiro Yoshida; Koji Fukuzawa; Asumi Takei; Gaku Kanda; Kaoru Takami; Hiroyuki Kumagai; Satoko Tanaka; Mitsuaki Itoh; Kimitake Imamura; Ryudo Fujiwara; Atsushi Suzuki; Ken-Ichi Hirata
Journal:  J Interv Card Electrophysiol       Date:  2011-04-20       Impact factor: 1.900