K W Lam1, J M Drake, R S Cobbold.
Abstract
Because normal cerebrospinal fluid (CSF) has almost no natural Doppler scatterers, patency testing of ventriculoperitoneal cerebrospinal fluid shunts (small silastic tubing with lumen diameter of approximately 1 mm draining excessive CSF from the brain) cannot be performed by Doppler ultrasound. We have developed a low-frequency bubble excitation system that generates microbubble scatterers in both distilled water and CSF. Doppler ultrasound can then be used for flow measurement in a ventriculoperitoneal shunt. By using low duty-cycle (approximately 10%), low-frequency (approximately 30 kHz), and low-amplitude (approximately 30 kPa) ultrasound, a population of microbubbles can be maintained for sufficiently long times (>10 min) for Doppler ultrasound measurement, although bubble initiation is inconsistent. The minimum pressure needed for bubble maintenance was found to decrease with increasing burst length and duty cycle. It has been possible to detect the presence of CSF shunt flow down to a mean flow rate of 3 mL/h (mean velocity approximately 0.6 mm/s). The bubble maintenance scheme developed satisfies the safety parameters specified by the American Institute of Ultrasound in Medicine (AIUM) and the US Food and Drug Administration (FDA). Results from both in vitro and in vivo (externalized shunts) experiments indicate the feasibility of this scheme for determining realistic CSF shunt flows, though some practical problems remain before the technique will be ready for clinical use.Entities:
Mesh:
Year: 1999 PMID: 10374981 DOI: 10.1016/s0301-5629(98)00164-1
Source DB: PubMed Journal: Ultrasound Med Biol ISSN: 0301-5629 Impact factor: 2.998