Respiratory resistance measured by an airflow perturbation device.

C G Lausted1, A T Johnson.   

Abstract

The airflow perturbation device (APD) is an instrument for the measurement of respiratory resistance. The APD is small, lightweight, fast and requires no special breathing manoeuvres. Airflow perturbation determines resistance by superimposing a periodic signal onto spontaneous breathing with a variable resistance device. Respiratory impedance is the ratio of magnitude of pressure perturbation to magnitude of flow perturbation, and respiratory resistance is the in-phase portion of respiratory impedance. The APD was tested to determine its responses to repeated resistance measurements and to changes in resistance. A mechanical model test showed that the APD could detect increased resistance levels, but overestimated resistance when resistance increased with flow. Tests with human subjects showed that the APD gave results consistent from day to day, was able to detect added resistances, and gave resistance values correlated with airway resistance values obtained by body plethysmography. Accelerometers placed on the chests of the subjects showed that perturbations extend to the chest surface. Thus, the APD must measure total respiratory resistance.

Entities:  

Mesh:

Year:  1999        PMID: 10374824     DOI: 10.1088/0967-3334/20/1/002

Source DB:  PubMed          Journal:  Physiol Meas        ISSN: 0967-3334            Impact factor:   2.833


  7 in total

1.  The Effect of Exercise on Respiratory Resistance in Athletes With and Without Paradoxical Vocal Fold Motion Disorder.

Authors:  Sally J K Gallena; Nancy Pearl Solomon; Arthur T Johnson; Jafar Vossoughi; Wei Tian
Journal:  Am J Speech Lang Pathol       Date:  2015-08       Impact factor: 2.408

2.  Validity of a new respiratory resistance measurement device to detect glottal area change.

Authors:  Sally J K Gallena; Wei Tian; Arthur T Johnson; Jafar Vossoughi; Stephen A Sarles; Nancy Pearl Solomon
Journal:  J Voice       Date:  2013-03-15       Impact factor: 2.009

3.  Contributions of signal analysis to the interpretation of spirometry.

Authors:  A Anogeianaki; N Negrev; G Ilonidis
Journal:  Hippokratia       Date:  2007-10       Impact factor: 0.471

4.  Measuring the mechanical input impedance of the respiratory system with breath-driven flow oscillations.

Authors:  Gregory S Roy; Nirav Daphtary; Olivia Johnson; Anne E Dixon; David A Kaminsky; Jason H T Bates
Journal:  J Appl Physiol (1985)       Date:  2021-02-11

5.  Decrease of resistance to air flow with nasal strips as measured with the airflow perturbation device.

Authors:  Lily S Wong; Arthur T Johnson
Journal:  Biomed Eng Online       Date:  2004-10-22       Impact factor: 2.819

6.  Testing limits to airflow perturbation device (APD) measurements.

Authors:  Erika R Lopresti; Arthur T Johnson; Frank C Koh; William H Scott; Shaya Jamshidi; Nischom K Silverman
Journal:  Biomed Eng Online       Date:  2008-10-31       Impact factor: 2.819

7.  Comparison of Respiratory Resistance Measurements Made with an Airflow Perturbation Device with Those from Impulse Oscillometry.

Authors:  J Pan; A Saltos; D Smith; A Johnson; J Vossoughi
Journal:  J Med Eng       Date:  2013-04-04
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.