Design of biplanar gradient coils for magnetic resonance imaging of the human torso and limbs.

G B Williams1, B J Fisher, C L Huang, T A Carpenter, L D Hall.   

Abstract

A method is described for design of gradient coils of unconventional geometry for MRI that is based on the superpositions of magnetic fields arising from individual current elements calculated by the Biot-Savart Law. Use of an optimization method based on a genetic algorithm enables a wide diversity in the shapes of coil that can be modeled. To exemplify this a two axis, biplanar gradient set is presented; this geometry offers good access for rectangular objects whilst holding the coils closer to the region of interest than is possible for cylindrical configurations. The inner dimensions of the gradient set were 40.0 x 24.4 x 40.0 cm and the gradient efficiencies were 0.3 and 0.4 mT m(-1) A(-1) in the z- and y- directions respectively over a 15 cm diameter region. Correction of signals arising from regions for which gradient linearity was not optimized was successful for the monotonic region within the set; the largest cuboid from which the MR signal could be processed to produce an undistorted image is of dimensions 36.3 x 17.2 x 24.4 cm.

Entities:  

Mesh:

Year:  1999        PMID: 10372528     DOI: 10.1016/s0730-725x(99)00012-0

Source DB:  PubMed          Journal:  Magn Reson Imaging        ISSN: 0730-725X            Impact factor:   2.546


  2 in total

1.  Superelliptical insert gradient coil with a field-modifying layer for breast imaging.

Authors:  Sung M Moon; K Craig Goodrich; J Rock Hadley; Seong-Eun Kim; Gengsheng L Zeng; Glen R Morrell; Matthew A McAlpine; Blaine A Chronik; Dennis L Parker
Journal:  Magn Reson Med       Date:  2010-10-11       Impact factor: 4.668

2.  Local planar gradients with order-of-magnitude strength and speed advantage.

Authors:  Bulent Aksel; Luca Marinelli; Bruce D Collick; Cornelius Von Morze; Paul A Bottomley; Christopher J Hardy
Journal:  Magn Reson Med       Date:  2007-07       Impact factor: 3.737

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.