Literature DB >> 10369350

Tagged MR imaging of intracranial aneurysm models.

H Isoda1, Y Kinosada, S Isogai, Y Takehara, T Ito.   

Abstract

We used tagged MR imaging to investigate the flow in two lateral and two terminal saccular intracranial aneurysm models of different neck sizes. Imaging was performed with a 1.5-T superconducting MR system using 2D fast spatial modulation of magnetization (SPAMM) sequences with an intersegmental delay of 25 milliseconds. The flow in the saccular aneurysm models varied with the shape and size of the neck: flow was faster in wider-necked aneurysms than in those with narrower necks.

Entities:  

Mesh:

Year:  1999        PMID: 10369350

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  1 in total

1.  Preliminary study of tagged MR image velocimetry in a replica of an intracranial aneurysm.

Authors:  Haruo Isoda; Shoichi Inagawa; Hiroyasu Takeda; Satoshi Isogai; Yasuo Takehara; Harumi Sakahara
Journal:  AJNR Am J Neuroradiol       Date:  2003-04       Impact factor: 3.825