High-level disinfection of gastrointestinal endoscopes: are current guidelines adequate?

B J Kovacs1, Y K Chen, J D Kettering, R M Aprecio, I Roy.   

Abstract

OBJECTIVE: For a germicide to obtain a high level disinfection (HLD) claim, FDA requires demonstration of a 6-log reduction of mycobacterial inoculum under worst case conditions. The purpose of this study was to assess the adequacy of current guidelines for high level disinfection of GI endoscopes using alkaline glutaraldehyde in simulated-use testing.
METHODS: Various gastrointestinal endoscopes were contaminated with Mycobacterium chelonae in 46 experiments. Quantitative cultures were obtained from each endoscope channel separately after each step: inoculation, standardized manual cleaning, immersion in 2% glutaraldehyde (Cidex) for 10, 20, or 45 min at room temperature, 70% isopropanol rinse, and drying.
RESULTS: Manual cleaning alone achieved a 4-log reduction. After 10 min of glutaraldehyde exposure, but before alcohol rinse, two of 10 experiments failed to achieve a 6-log reduction. However, after alcohol rinse, all 10 experiments achieved HLD. After 20 min of glutaraldehyde exposure, but before alcohol rinse, one of 18 experiments failed to achieve a 6-log reduction. After alcohol rinse, all 18 experiments achieved HLD. After 45 min of glutaraldehyde exposure, but before alcohol rinse, one of 18 experiments failed to achieve a 6-log reduction. After alcohol rinse, all 18 experiments achieved HLD. Thus, if the entire reprocessing protocol including manual cleaning, glutaraldehyde exposure, alcohol rinse, and drying was taken into account, the required 6-log reduction of mycobacteria was achieved with a minimum of 10 min of glutaraldehyde exposure at room temperature.
CONCLUSIONS: Current guidelines for high level disinfection using glutaraldehyde are appropriate. Alcohol rinse is a valuable adjunctive step for drying and for its bactericidal effects.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10364023     DOI: 10.1111/j.1572-0241.1999.01142.x

Source DB:  PubMed          Journal:  Am J Gastroenterol        ISSN: 0002-9270            Impact factor:   10.864


  4 in total

1.  Current GI endoscope disinfection and QA practices.

Authors:  Frank M Moses; Jennifer S Lee
Journal:  Dig Dis Sci       Date:  2004 Nov-Dec       Impact factor: 3.199

Review 2.  Is peracetic acid suitable for the cleaning step of reprocessing flexible endoscopes?

Authors:  Günter Kampf; Patricia M Fliss; Heike Martiny
Journal:  World J Gastrointest Endosc       Date:  2014-09-16

Review 3.  Preventing nosocomial infections from gastrointestinal endoscopy.

Authors:  D Franchi; A Bahrani; J F Ober; M B Edmond
Journal:  Curr Gastroenterol Rep       Date:  2000-08

4.  Effect of Disinfectants on Mechanical Properties of Orthodontic Acrylics.

Authors:  Tobias Bensel; Jens J Bock; Anne Kebernik; Christin Arnold; Sonia Mansour; Arne F Boeckler
Journal:  Int J Biomater       Date:  2019-04-24
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.