Do the components of heat and moisture exchanger filters affect their humidifying efficacy and the incidence of nosocomial pneumonia?

L Thomachot1, R Vialet, S Arnaud, B Barberon, A Michel-Nguyen, C Martin.   

Abstract

OBJECTIVES: To compare the efficiency of two heat and moisture exchange filters (HMEFs) of different compositions of the humidifying capacity and the rate of bronchial colonization and ventilator-associated pneumonia in patients in the intensive care unit (ICU).
DESIGN: Prospective, randomized study.
SETTING: ICU of a university hospital. PATIENTS: All patients who required mechanical ventilation for 24 hrs or more during the study period.
INTERVENTIONS: At admission to the ICU, patients were randomly assigned to one of two groups. In one group, the patients were ventilated with Humid-Vent Filter Light HMEF. The condensation surface was made of paper impregnated with CaCl2. The filter membrane was made of polypropylene. In the other group, the patients were ventilated with the Clear ThermAl HMEF (Intersurgical, France). The condensation surface was made of plastic foam impregnated with AlCl2. The filter membrane was made of two polymer fibers (modacrylic and polypropylene). In both groups, HMEFs were changed daily.
MEASUREMENTS AND MAIN RESULTS: Seventy-seven patients were ventilated for 19+/-7 days with the Humid-Vent Filter Light HMEF and 63 patients for 17+/-6 days with the Clear ThermAl HMEF. Patients ventilated with the Humid-Vent Filter Light underwent 8.7+/-3.7 tracheal aspirations and 1.2+/-2.0 instillations per day and those with the Clear ThermAl, 8.2+/-3.9 and 1.5+/-2.4 per day, respectively (NS). The abundance of tracheal secretions and the presence of blood and viscosity, as evaluated by semiquantitative scales, were similar in both groups. One episode of tracheal tube occlusion was observed with the Humid-Vent Filter Light HMEF and none with the other HMEF (NS). Tracheal colonization was observed at a rate of 91% with the Humid-Vent Filter Light and 97% with the Clear ThermAl (NS). The rate of ventilator-associated pneumonia was similar in both groups (35%). Bacteria responsible for tracheal colonization and pneumonia were similar in both groups.
CONCLUSIONS: Despite differences in their components, the two HMEFs that were tested achieved similar performances in terms of humidification and heating of inspired gases. Only one episode of endotracheal tube occlusion was detected, and very few patients (three in each group) had to be switched to an active heated humidifier. No difference was observed either in the rate of tracheal colonization or of ventilator-associated pneumonia. These data show that the Humid-Vent Filter Light and the Clear ThermAl HMEFs are suited for use with ICU patients.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10362414     DOI: 10.1097/00003246-199905000-00026

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  6 in total

Review 1.  Efficacy of heat and moisture exchangers in preventing ventilator-associated pneumonia: meta-analysis of randomized controlled trials.

Authors:  Axel Kola; Tim Eckmanns; Petra Gastmeier
Journal:  Intensive Care Med       Date:  2004-09-11       Impact factor: 17.440

2.  Under-humidification and over-humidification during moderate induced hypothermia with usual devices.

Authors:  François Lellouche; Siham Qader; Solenne Taille; Aissam Lyazidi; Laurent Brochard
Journal:  Intensive Care Med       Date:  2006-05-23       Impact factor: 17.440

3.  Impact of selective digestive decontamination on respiratory tract Candida among patients with suspected ventilator-associated pneumonia. A meta-analysis.

Authors:  J C Hurley
Journal:  Eur J Clin Microbiol Infect Dis       Date:  2016-04-26       Impact factor: 3.267

4.  A new device for 100 per cent humidification of inspired air.

Authors:  A Larsson; A Gustafsson; L Svanborg
Journal:  Crit Care       Date:  2000-01-24       Impact factor: 9.097

5.  Unusually High Incidences of Staphylococcus aureus Infection within Studies of Ventilator Associated Pneumonia Prevention Using Topical Antibiotics: Benchmarking the Evidence Base.

Authors:  James C Hurley
Journal:  Microorganisms       Date:  2018-01-04

6.  Influence of a heat and moisture exchanger with a microbiological filter on measurements of maximal respiratory pressures and vital capacity in patients with COPD.

Authors:  Jeanette Janaina Jaber Lucato; Renata Cléia Claudino Barbosa; Patricia Salerno de Almeida Picanço; Thiago Marraccini Nogueira da Cunha; Renato Fraga Righetti
Journal:  J Bras Pneumol       Date:  2019-12-13       Impact factor: 2.624

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.