PURPOSE: To develop a noninvasive technique to visualize and measure blood flow in the iris and ciliary body.
METHODS: Echo data from 50-MHz ultrasound scans of the iris and ciliary body of rabbits were digitized using a new "swept scan" modality. The method makes use of spatial oversampling to identify regions with scatterers whose range changes with time. The data allowed construction of high-resolution B-mode images with embedded flow information. Pulsatility over the cardiac cycle was evaluated by sending a series of pulses along a single line of sight containing a vessel of interest. Local blood flow and changes over the cardiac cycle before and after application of atropine were quantified.
RESULTS: Flow was identified in the radial vessels and major arterial circle of the iris. Vessels with lumens as small as 40 microm in diameter and flow velocities as low as 0.6 mm/sec were measured. Change in blood velocity over the cardiac cycle was determined to be approximately 27%. Peak systolic velocity after administration of topical atropine increased by 72%.
CONCLUSIONS: This technique allowed visualization of flow using the same type of very-high-frequency transducer now widely used for imaging the anterior segment. The technique can also be used at lower frequencies for more posterior tissues with similar improvement of resolution over Doppler. The ability to examine flow in the anterior segment of the eye offers a new tool for study of glaucoma, hypotony, tumors, and other disorders.

Entities:  

Mesh:

Year:  1999        PMID: 10359319

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  6 in total

1.  Explaining The Current Role Of High Frequency Ultrasound In Ophthalmic Diagnosis (Ophthalmic Ultrasound).

Authors:  D Jackson Coleman; Ronald H Silverman; Mark J Rondeau; Harriet O Lloyd; Suzanne Daly
Journal:  Expert Rev Ophthalmol       Date:  2006-10-01

2.  Measurement of choroidal perfusion and thickness following systemic sildenafil (Viagra(®) ).

Authors:  David Y Kim; Ronald H Silverman; Robison Vernon Paul Chan; Aziz A Khanifar; Mark Rondeau; Harriet Lloyd; Peter Schlegel; Donald Jackson Coleman
Journal:  Acta Ophthalmol       Date:  2012-09-13       Impact factor: 3.761

3.  Enhanced piezoelectric performance of composite sol-gel thick films evaluated using piezoresponse force microscopy.

Authors:  Yuanming Liu; Kwok Ho Lam; K Kirk Shung; Jiangyu Li; Qifa Zhou
Journal:  J Appl Phys       Date:  2013-05-08       Impact factor: 2.546

Review 4.  Dual-Mode Tumor Imaging Using Probes That Are Responsive to Hypoxia-Induced Pathological Conditions.

Authors:  S A Amali S Subasinghe; Robia G Pautler; Md Abul Hassan Samee; Jason T Yustein; Matthew J Allen
Journal:  Biosensors (Basel)       Date:  2022-06-30

Review 5.  High-resolution ultrasound imaging of the eye - a review.

Authors:  Ronald H Silverman
Journal:  Clin Exp Ophthalmol       Date:  2008-12-09       Impact factor: 4.207

6.  Blood velocity measurement in the posterior segment of the rabbit eye using combined spectral Doppler and power Doppler ultrasound.

Authors:  Walid Abdallah; Amani Fawzi; Hitenkumar Patel; Grant Dagliyan; Naoki Matsuoka; Edward Grant; Mark Humayun
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-10-03       Impact factor: 3.117

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.