Literature DB >> 10357488

Polytetrafluoroethylene-encapsulated stent-grafts: use in experimental abdominal aortic aneurysm.

A E Benson1, J C Palmaz, F O Tio, E A Sprague, C E Encarnacion, S C Josephs.   

Abstract

PURPOSE: To evaluate expanded polytetrafluoroethylene (ePTFE) encapsulated stents for the treatment of aortic aneurysms with emphasis on the blood and tissue-material interactions.
MATERIALS AND METHODS: Experimental aortic aneurysms were created in dogs by enlarging the aortic lumen with an abdominal fascial patch. Twenty animals underwent endoluminal repair after allowing the surgically created aneurysm to heal for 2 months prior to transluminal aneurysmal exclusion. The device used consisted of an 8-cm-long ePTFE encapsulated stent graft. The animals were killed in groups at 1 week and at 1, 2.25, 6, and 12 months. Specimens were processed for histologic and luminal surface studies.
RESULTS: Before the animals were killed, aortography demonstrated two thrombosed aortae in the 6-month group and two endoleaks in the 12-month group. Endothelialized neointima extended into the proximal and distal portions of the prosthetic lumen, with minimal cell coverage in the center of the graft. The overall percent surface area covered by endothelialized neointima was 22% +/- 6% at 6 months and 18% +/- 10% by 1 year (P = .75). Histologic examination demonstrated minimal tissue penetration into the ePTFE.
CONCLUSION: Transluminal exclusion of abdominal aortic aneurysms by encapsulated stent-graft is easily accomplished. With this device, tissue coverage and penetration of the stent graft is limited and does not tend to increase with time.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10357488     DOI: 10.1016/s1051-0443(99)70091-6

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  2 in total

1.  Tissue responses to endovascular stent grafts for saccular abdominal aortic aneurysms in a canine model.

Authors:  Hyun Beom Kim; Young Ho Choi; Young Ho So; Seung-Kee Min; Hyo-Cheol Kim; Young Il Kim; Jae Hyung Park; Jin Wook Chung
Journal:  J Korean Med Sci       Date:  2012-10-02       Impact factor: 2.153

2.  Tissue responses to stent grafts with endo-exo-skeleton for saccular abdominal aortic aneurysms in a canine model.

Authors:  Young Il Kim; Young Ho Choi; Jin Wook Chung; Hyo-Cheol Kim; Young Ho So; Hyun Beom Kim; Seung-Kee Min; Jae Hyung Park
Journal:  Korean J Radiol       Date:  2014-09-12       Impact factor: 3.500

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.