Literature DB >> 10357487

Relationship between blood flow, thrombus, and neointima in stents.

G M Richter1, J C Palmaz, G Noeldge, F Tio.   

Abstract

PURPOSE: To establish a relationship between flow, acute thrombus formation, and late intimal hyperplasia in arterial stents.
MATERIALS AND METHODS: To compare short-term stent patency in the canine femoral artery with normal flow to that in the opposite femoral artery with restricted flow, 24 dogs were subdivided in four groups: groups 1 (no intravenous heparin) and 2 (intravenous heparin) had unilateral flow restriction by surgically created stenosis, downstream of a Palmaz stent. Group 3 (no intravenous heparin) and 4 (intravenous heparin) had sham surgical exposure of the corresponding arterial segment, without flow restriction. Thrombocyte activity over the stent segment was evaluated for 3 hours after stent placement with nuclear scanning, after administration of indium-111-labeled platelets. To evaluate long-term stent patency in relationship to arterial flow, 14 additional dogs were subjected to long-term observation. Matched, symmetrically implanted femoral stents with normal and restricted flow, were explanted at 1, 12, and 24 weeks for histologic analysis and comparative measurement of neointimal thickness. Angiographic studies were performed before and after nuclear scanning in the short-term study group and before explant in the long-term animal group.
RESULTS: In the short-term, groups 2 and 4 showed neither increased platelet uptake nor angiographically demonstrable thrombus. Group 1 had increased platelet uptake and occlusive or subocclusive angiographic thrombus. Group 3 had increased platelet uptake and angiographic thrombus in one instance. In the long-term, stents with flow restriction had significantly greater neointimal formation in comparison with unrestricted stents. Histologic studies suggested that the stent neointima resulted from progressive replacement of stent thrombus.
CONCLUSION: Regardless of flow condition, intravenous heparinization is necessary to prevent thrombus formation in the stent lumen. Within the experimental parameters of this study, low flow and absent heparinization consistently lead to stent thrombosis. Stent implantation under low flow is associated with increased neointima formation. It is not known whether this is preventable by antithrombotic medication.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10357487     DOI: 10.1016/s1051-0443(99)70090-4

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  7 in total

1.  Polyester vascular patches acquire arterial or venous identity depending on their environment.

Authors:  Hualong Bai; Haidi Hu; Jianming Guo; Maryam Ige; Tun Wang; Toshihiko Isaji; Tambudzai Kudze; Haiyang Liu; Bogdan Yatsula; Takuya Hashimoto; Ying Xing; Alan Dardik
Journal:  J Biomed Mater Res A       Date:  2017-09-26       Impact factor: 4.396

2.  "Full metal jacket" with direct stenting of complete chronic occlusions of the superficial femoral artery.

Authors:  D Laganà; G Carrafiello; M Barresi; D Lumia; M Dizonno; F A Vizzari; F Fontana; M Mangini; P Castelli; C Fugazzola
Journal:  Radiol Med       Date:  2011-01-12       Impact factor: 3.469

3.  Abdominal aortic aneurysm: Treatment options, image visualizations and follow-up procedures.

Authors:  Zhong-Hua Sun
Journal:  J Geriatr Cardiol       Date:  2012-03       Impact factor: 3.327

4.  Pericardial patch venoplasty heals via attraction of venous progenitor cells.

Authors:  Hualong Bai; Mo Wang; Trenton R Foster; Haidi Hu; Hao He; Takuya Hashimoto; Jesse J Hanisch; Jeans M Santana; Ying Xing; Alan Dardik
Journal:  Physiol Rep       Date:  2016-06

5.  Continuously Grooved Stent Struts for Enhanced Endothelial Cell Seeding.

Authors:  Marja Ter Meer; Willeke F Daamen; Yvonne L Hoogeveen; Gijs J F van Son; Jeremy E Schaffer; J Adam van der Vliet; Leo J Schultze Kool; Lambertus P van den Heuvel
Journal:  Cardiovasc Intervent Radiol       Date:  2017-05-03       Impact factor: 2.740

6.  Fenestrated stent graft repair of abdominal aortic aneurysm: hemodynamic analysis of the effect of fenestrated stents on the renal arteries.

Authors:  Zhonghua Sun; Thanapong Chaichana
Journal:  Korean J Radiol       Date:  2009-12-28       Impact factor: 3.500

7.  Self-expandable nitinol stent placement in homocysteinemic porcine aorta.

Authors:  Luís Henrique Gil França; Adamastor Humberto Pereira; Sílvio César Perini
Journal:  Clinics (Sao Paulo)       Date:  2008-04       Impact factor: 2.365

  7 in total