Accuracy of radial artery blood pressure determination with the Vasotrac.

K G Belani1, J J Buckley, M O Poliac.   

Abstract

PURPOSE: To evaluate the accuracy of a new non-invasive method (Vasotrac) to measure blood pressure (BP) with accompanying arterial wave-form and pulse-rate display when compared with BP and waveform measured invasively.
METHODS: Healthy volunteers (n=53) served as subjects for the study. Blood pressures and waveforms obtained via a radial artery catheter (IABP) were compared with non-invasive measurements obtained every 12-15 beats by the new system (Vasotrac BP) from the opposite radial artery. In a sub-group of volunteers (n=11), BP was acutely increased and decreased with isoproterenol, phenylephrine and sodium nitroprusside. Data were analyzed by determining correlation and agreement between the two systems of measurement. Waveforms obtained by the two systems were qualitatively examined.
RESULTS: Non-invasive BP measured every 12-15 beats by the Vasotrac correlated with IABP (systolic r2 = 0.89; diastolic r2 = 0.88; mean r2 = 0.94). The actual values obtained by the two methods agreed closely with > 90% of readings being within 2SDs when plotted by the Bland Altman method. This was also true during vasoactive drug infusion when BP changed acutely and swiftly. During this dynamic period, Vasotrac BP accurately tracked the changes in IABP with correlations (systolic r2 = 0.82; diastolic r2 = 0.89; mean r2 = 0.95) and close agreement (> 90% of readings were within 2 SDs in the Bland Altman plot). Waveforms displayed by the two systems were qualitatively very similar. Pulse rates measured by the two systems were identical.
CONCLUSIONS: The Vasotrac system displayed an arterial waveform which was similar to that obtained directly and measured BP and pulse rate accurately. It should be a convenient device to measure BP continually in a non-invasive fashion.

Entities:  

Mesh:

Year:  1999        PMID: 10349931     DOI: 10.1007/BF03012951

Source DB:  PubMed          Journal:  Can J Anaesth        ISSN: 0832-610X            Impact factor:   5.063


  6 in total

1.  Innovative continuous non-invasive cuffless blood pressure monitoring based on photoplethysmography technology.

Authors:  Juan C Ruiz-Rodríguez; Adolf Ruiz-Sanmartín; Vicent Ribas; Jesús Caballero; Alejandra García-Roche; Jordi Riera; Xavier Nuvials; Miriam de Nadal; Oriol de Sola-Morales; Joaquim Serra; Jordi Rello
Journal:  Intensive Care Med       Date:  2013-06-06       Impact factor: 17.440

2.  Heart Rate Variability Moderates the Association Between Separation-Related Psychological Distress and Blood Pressure Reactivity Over Time.

Authors:  Kyle J Bourassa; Karen Hasselmo; David A Sbarra
Journal:  Psychol Sci       Date:  2016-06-14

Review 3.  Detecting vital signs with wearable wireless sensors.

Authors:  Tuba Yilmaz; Robert Foster; Yang Hao
Journal:  Sensors (Basel)       Date:  2010-12-02       Impact factor: 3.576

Review 4.  Current Status and Prospects of Health-Related Sensing Technology in Wearable Devices.

Authors:  Jaegeol Cho
Journal:  J Healthc Eng       Date:  2019-06-16       Impact factor: 2.682

5.  Non-invasive monitoring of blood pressure using the Philips Intellivue MP50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia.

Authors:  Xianghu Meng; Guanghui Zang; Longchang Fan; Lei Zheng; Jinzhen Dai; Xueren Wang; Wei Xia; Jihong Liu; Chuanhan Zhang
Journal:  Exp Ther Med       Date:  2013-05-17       Impact factor: 2.447

6.  Double-leg isometric exercise training in older men.

Authors:  Anthony W Baross; Jonathan D Wiles; Ian L Swaine
Journal:  Open Access J Sports Med       Date:  2013-01-30
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.