M Obayashi1, T Kosugi, J Yamazaki, Y Matsumoto, M Fukuoka, M Matsumoto.
Abstract
An optimized method for the determination of flecainide in serum is presented. Extraction using a solid-phase C18 column and chromatography on a stabilized fluorocarbon-bonded silica gel column effectively separate flecainide from an internal standard (a positional isomer of flecainide). The HPLC apparatus and conditions were as follows: analytical column, Fluofix 120N; sample solvent, 20 microl; column temperature, 40 degrees C; detector, Shimadzu RF-5000 fluorescence spectrophotometer (excitation wavelength = 300 nm, emission wavelength = 370 nm); mobile phase, 0.06% phosphoric acid containing 0.1% tetra-n-butyl ammonium bromide-acetonitrile (75:25, v/v); flow-rate, 1.0 ml/min. The standard curves for flecainide were linear in the concentration range examined (10-2000 ng/ml). The regression equation was y = 0.08+0.0078x (r = 0.9998). The minimum detectable amount of flecainide was approximately 5 ng/ml. In the within-day study, the precision coefficients of variation were 2.66, 2.18, 2.54, 2.72, 2.88, 2.24, and 3.29% for the 10, 50, 100, 200, 500, 1000, and 1500 ng/ml standards, respectively. The absolute recovery rates of flecainide at each concentrations were 94-100%. The method described provides analytical sensitivity, specificity and reproducibility suitable for both biomedical research and therapeutic drug monitoring.Entities:
Mesh:
Substances:
Year: 1999 PMID: 10348189 DOI: 10.1016/s0378-4347(98)00578-7
Source DB: PubMed Journal: J Chromatogr B Biomed Sci Appl ISSN: 1387-2273