Literature DB >> 10341842

Stents covered by an autologous arterial graft in porcine coronary arteries: feasibility, vascular injury and effect on neointimal hyperplasia.

C Stefanadis1, K Toutouzas, E Tsiamis, C Vlachopoulos, S Vaina, D Tsekoura, L Haldi, E Stefanadi, M Gravanis, P Toutouzas.   

Abstract

OBJECTIVE: The use of stents has improved results after balloon coronary angioplasty. Several materials have been proposed for covering the metallic surface of the stent to reduce the rate of subacute thrombosis and restenosis. In our institution, an autologous arterial graft was used for covering the external surface of a conventional stent. The angiographic and histological response in a porcine coronary artery model was investigated.
METHODS: An autologous arterial graft was removed from the femoral artery and carefully prepared. Subsequently, a conventional stent was covered externally by the arterial graft. Twenty-two covered stents and 22 uncovered regular stents were implanted alternatively in the coronary arteries of 22 pigs. One animal died immediately after the procedure, due to thrombus formation in the uncovered stent. Six animals were sacrificed at seven days and the remaining animals were sacrificed at two months. Before the sacrifice, coronary angiography was performed in all animals.
RESULTS: Thrombosis was detected in two control segments and in one covered stented segment. After seven days, the luminal surface of the covered stents was covered by a new endothelial layer in contrast to partial endothelial cell appearance in the control group. The angiographic parameters were similar between the two groups. Histologically, the covered stents were associated with less vascular injury compared to uncovered stents. In covered stents a trend towards reduction of maximal intimal hyperplasia was detected (covered: 116.6 +/- 47.75 vs uncovered: 150.25 +/- 46.81 microns, p = 0.08); also the thickness of the arterial media was reduced (covered: 21.34 +/- 10.28 vs uncovered: 102.63 +/- 18.71 microns, p = 0.02). The luminal and vessel areas were similar in the two groups.
CONCLUSIONS: The preparation and implantation of the autologous arterial graft-covered stent is technically safe and feasible. This type of covered stent results in accelerated endothelialization, less vascular injury, thinning of the arterial media and a trend to reduce the intimal hyperplasia in normal coronary arteries.

Entities:  

Mesh:

Year:  1999        PMID: 10341842     DOI: 10.1016/s0008-6363(98)00211-9

Source DB:  PubMed          Journal:  Cardiovasc Res        ISSN: 0008-6363            Impact factor:   10.787


  2 in total

1.  A new stent graft. With thin walled controlled micropored polymer covering.

Authors:  S Nishi; Y Nakayama; H Ueda; M Ishikawa; T Matsuda
Journal:  Interv Neuroradiol       Date:  2001-05-15       Impact factor: 1.610

2.  Clinical Outcomes of Self-Made Polyurethane-Covered Stent Implantation for the Treatment of Coronary Artery Perforations.

Authors:  Xiaoyue Song; Qing Qin; Shufu Chang; Rende Xu; Mingqiang Fu; Hao Lu; Lei Ge; Juying Qian; Jianying Ma; Junbo Ge
Journal:  J Interv Cardiol       Date:  2021-05-17       Impact factor: 2.279

  2 in total