Accuracy of MR phase mapping for temperature monitoring during interstitial laser coagulation (ILC) in the liver at rest and simulated respiration.

J Heisterkamp1, N A Matheijssen, R van Hillegersberg, J J van Vaals, J S Laméris, J Stoker, J N Jzermans.   

Abstract

The chemical shift or proton-resonance frequency (phase mapping) can be used to measure temperature changes. As a subtraction technique, it requires scans at exactly the same location, making it prone to respiration-induced artifacts. The accuracy of magnetic resonance (MR) phase mapping for temperature monitoring of interstitial laser coagulation (ILC) was therefore investigated in two ex vivo models with simulated respiration. MR temperatures were calibrated to interstitially measured temperature. Gradual cooling of a homogenous medium (gel) was monitored for four starting temperatures (room temperature, 40 degrees C, 50 degrees C, and 60 degrees C) during 30 min. Temperature increases were measured during ILC in ex vivo porcine liver with Nd:YAG for 6 min with 5 Watt. Experiments were performed at rest and with simulated respiratory motion (both n = 5). In liver, accuracy did not decrease with respiration simulation (P = 0.32), whereas a significant decline was found in the gel model (P = 0.002). In all experiments a small drift over time was observed between temperature determined with MR and thermoprobes. Correction for temperature-independent phase-shift at a reference location did not enhance agreement. Temperatures could be determined correctly by MR in the moving liver within a range of +/-3.5 degrees C after 6 min of laser application (95% confidence interval), justifying further pre-clinical studies.

Mesh:

Year:  1999        PMID: 10332874     DOI: 10.1002/(sici)1522-2594(199905)41:5<919::aid-mrm10>3.0.co;2-4

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  2 in total

1.  The influence of Nd:YAG laser irradiation on Fluoroptic® temperature measurement: an experimental evaluation.

Authors:  Frank Hübner; Babak Bazrafshan; Jörg Roland; Antje Kickhefel; Thomas Josef Vogl
Journal:  Lasers Med Sci       Date:  2012-04-04       Impact factor: 3.161

Review 2.  MR fluoroscopy in vascular and cardiac interventions (review).

Authors:  Maythem Saeed; Steve W Hetts; Joey English; Mark Wilson
Journal:  Int J Cardiovasc Imaging       Date:  2011-02-26       Impact factor: 2.357

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.