Literature DB >> 10332783

Corrosion on an internal spinal fixator system.

U Vieweg1, D van Roost, H K Wolf, C A Schyma, J Schramm.   

Abstract

STUDY
DESIGN: Thirteen spinal fixators with 26 stabilization bridges and 52 pedicle screws and a mean length of implantation of 10 months were prospectively examined for corrosion.
OBJECTIVES: To determine the type of corrosion and the correlation between the construction of the spinal fixator and the type of corrosion. SUMMARY OF BACKGROUND DATA: Evidence of fretting and crevice corrosion is seen in many stainless steel implants in retrieval studies. Such reactions have not been described in the literature on spinal fixator systems.
METHODS: Macroscopic and microscopic alterations in the adjacent tissue were examined, and the corrosive alterations were documented photographically using stereoscopic optical light microscopy. The chemical composition of the implants was determined spectrographically. Microradiography and x-ray fluorescence analysis of the soft tissue were performed.
RESULTS: At surgery, tissue discoloration was found in four cases. Histologic examination showed extensive fibrosis, foreign body reaction and inflammation associated with a small number of metal particles, indicating metallosis in five cases. Corresponding particles were detected by microradiography. Corrosion was found on 13 telescopic rods and on two pedicle screws. The alterations on the telescopic rods could be interpreted as crevice corrosion and the alterations in the pedicle screws as fretting corrosion. The two monobloc fixator bridges did not show signs of corrosion. In these implants, the neighboring tissue was macroscopically inconspicuous, and histologic examination showed minimal fibrosis or presence of metal particles. Spectrographic examination of the spinal fixators showed no structural imperfection.
CONCLUSIONS: The construction constraints of a spinal fixator make it prone to corrosion. New spinal implants should be examined not only in vitro but also in vivo to ascertain whether corrosion and adjacent tissue reaction occur. Corrosion is one reason to explant the internal fixation system after fusion of the spinal fracture.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10332783     DOI: 10.1097/00007632-199905150-00004

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  8 in total

1.  Osteomyelitis associated with an orthopedic implant.

Authors:  Sheila C Rahal; Maria J Mamprim; Evelyn H G Caporali; Carlos R Teixeira
Journal:  Can Vet J       Date:  2003-07       Impact factor: 1.008

2.  Metal levels in corrosion of spinal implants.

Authors:  Javier del Rio; Jose Beguiristain; Julio Duart
Journal:  Eur Spine J       Date:  2007-01-26       Impact factor: 3.134

3.  Comparative analysis between shape memory alloy-based correction and traditional correction technique in pedicle screws constructs for treating severe scoliosis.

Authors:  Yan Wang; Guoquan Zheng; Xuesong Zhang; Yonggang Zhang; Songhua Xiao; Zheng Wang
Journal:  Eur Spine J       Date:  2009-11-10       Impact factor: 3.134

Review 4.  Metallosis after traumatic loosening of Bryan cervical disc arthroplasty: a case report and literature review.

Authors:  Chih-Chan Yang; Chien-Lun Tang; Chung-Yuh Tzeng; Hsi-Kai Tsou
Journal:  Eur Spine J       Date:  2017-11-21       Impact factor: 3.134

5.  The outcome of pedicle screw instrumentation removal for ongoing low back pain following posterolateral lumbar fusion.

Authors:  Mario G Zotti; Oscar P Brumby-Rendell; Ben McDonald; Tom Fisher; Christovalantis Tsimiklis; Wai Weng Yoon; Orso L Osti
Journal:  J Spine Surg       Date:  2015-12

6.  Temporary use of shape memory spinal rod in the treatment of scoliosis.

Authors:  Yan Wang; Guoquan Zheng; Xuesong Zhang; Yonggang Zhang; Songhua Xiao; Zheng Wang
Journal:  Eur Spine J       Date:  2010-07-14       Impact factor: 3.134

7.  Metallosis: A Complication in the Guided Growing Rod System Used in Treatment of Scoliosis.

Authors:  Jae Hyuk Yang; Chang Hwa Ham; Yeok Gu Hwang; Seung Woo Suh
Journal:  Indian J Orthop       Date:  2017 Nov-Dec       Impact factor: 1.251

Review 8.  Holistic Approach to the Diagnosis and Treatment of Patients with Tumor Metastases to the Spine.

Authors:  Hanna Nowak; Dominika Maria Szwacka; Monika Pater; Wojciech Krzysztof Mrugalski; Michał Grzegorz Milczarek; Magdalena Staniszewska; Roman Jankowski; Anna-Maria Barciszewska
Journal:  Cancers (Basel)       Date:  2022-07-18       Impact factor: 6.575

  8 in total