Mapping retinal fluorescein leakage with confocal scanning laser fluorometry of the human vitreous.

C L Lobo1, R C Bernardes, F J Santos, J G Cunha-Vaz.   

Abstract

OBJECTIVE: To demonstrate an objective, quantitative, and sensitive method of mapping retinal fluorescein leakage into the vitreous while simultaneously imaging the retina.
METHODS: A prototype Zeiss confocal scanning laser ophthalmoscope was modified to obtain fluorometric measurements from 18 optical planes across the retina and cortical vitreous, separated from each other by 150 microm, and parallel to the retinal surface. After intravenous administration of fluorescein, an axial graphic of equivalent fluorescein concentration in the vitreous may be obtained from any region of interest. After correcting for fluorescence levels in the retina and choroid and plasma levels of free fluorescein, permeability values of the blood-retinal barrier to fluorescein were obtained from 1512 regions measuring 75 x 75 microm, from a total 3150 x 2700-microm area of the fundus, generating a detailed map of retinal fluorescein leakage. The method was assessed in vitro and in 7 healthy subjects who underwent scans during separate visits. Depth resolution and influence of chorioretinal fluorescence were further tested in 2 patients with multiple drusen and in 2 eyes after vitrectomy. Fourteen eyes from 7 patients with diabetes and nonproliferative retinopathy were also examined. Lateral resolution was tested in 3 diabetic eyes that underwent focal photocoagulation. Four eyes from 2 patients with diabetes and minimal retinopathy were examined at 3-month intervals. All eyes examined had less than 2 diopters of astigmatism.
RESULTS: Characteristics of the modified confocal scanning laser fluorometer included a lower limit of detection equal to 0.40 Eq ng/mL and depth precision of +/-15 microm. Values for the blood-retinal barrier permeability index in healthy subjects, measured 30 minutes after a single intravenous pulse of fluorescein (14 mg/kg), ranged from 1.3 +/- 0.4 x 10(-6) cm/s over the foveal avascular zone to 2.2 +/- 0.6 x 10(-6) cm/s over vessels in the retina. Diabetic eyes with retinopathy showed higher values, ranging from 1.4 to 15.0 x 10(-6) cm/s. Vitrectomized eyes and eyes with multiple drusen showed the validity of the correction algorithm demonstrating that measurements of fluorescence in the vitreous are not influenced by the chorioretinal fluorescence level. Argon laser photocoagulation burns placed in the diabetic retina demonstrated a lateral resolution on the order of 75 to 100 microm. Intravisit and intervisit reproducibility was +/-10.2% and +/-13%, respectively.
CONCLUSIONS: This new method measures localized alterations of the blood-retinal barrier and allows for direct correlation with retinal anatomy. Its most interesting feature is the ability to map retinal fluorescein leakage while simultaneously imaging the retina. This capability is expected to improve our understanding and management of retinal disease.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10326960     DOI: 10.1001/archopht.117.5.631

Source DB:  PubMed          Journal:  Arch Ophthalmol        ISSN: 0003-9950


  6 in total

1.  Agreement between clinical estimation and a new quantitative analysis by Photoshop software in fundus and angiographic image variables.

Authors:  Alireza Ramezani; Hamid Ahmadieh; Mohsen Azarmina; Masoud Soheilian; Mohammad H Dehghan; Mohammad R Mohebbi
Journal:  Int Ophthalmol       Date:  2008-09-19       Impact factor: 2.031

Review 2.  The significance of vascular and neural apoptosis to the pathology of diabetic retinopathy.

Authors:  Alistair J Barber; Thomas W Gardner; Steven F Abcouwer
Journal:  Invest Ophthalmol Vis Sci       Date:  2011-02-28       Impact factor: 4.799

3.  Novel fluorescein angiography-based computer-aided algorithm for assessment of retinal vessel permeability.

Authors:  Yonatan Serlin; Geva Tal; Yoash Chassidim; Yisrael Parmet; Oren Tomkins; Boris Knyazer; Alon Friedman; Jaime Levy
Journal:  PLoS One       Date:  2013-04-23       Impact factor: 3.240

Review 4.  Validation of structural and functional lesions of diabetic retinopathy in mice.

Authors:  T S Kern; J Tang; B A Berkowitz
Journal:  Mol Vis       Date:  2010-10-19       Impact factor: 2.367

5.  Overlay of conventional angiographic and en-face OCT images enhances their interpretation.

Authors:  Mirjam E J van Velthoven; Koos de Vos; Frank D Verbraak; Chris W Pool; Marc D de Smet
Journal:  BMC Ophthalmol       Date:  2005-06-13       Impact factor: 2.209

6.  Imaging Features of Vessels and Leakage Patterns Predict Extended Interval Aflibercept Dosing Using Ultra-Widefield Angiography in Retinal Vascular Disease: Findings From the PERMEATE Study.

Authors:  Azam Moosavi; Natalia Figueiredo; Prateek Prasanna; Sunil K Srivastava; Sumit Sharma; Anant Madabhushi; Justis P Ehlers
Journal:  IEEE Trans Biomed Eng       Date:  2021-05-21       Impact factor: 4.756

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.