Literature DB >> 10321538

The effect of fluorescein on the accuracy of blood glucose determination by glucose meters.

R J Antcliff1, C E Hugkulstone.   

Abstract

PURPOSE: To investigate whether intravenous fluorescein significantly affected the accuracy of a glucose meter.
METHODS: Thirty-four consecutive patients undergoing fluorescein angiography were recruited. Venous blood was removed immediately before and 1 hour after injection of 5 mls of 20% Sodium Fluorescein solution (1 mg). Whole blood glucose was measured with a glucose meter and plasma glucose levels were measured by the Department of Biochemistry for both samples. The results were analysed using Wilcoxon matched pairs test.
RESULTS: Mean glucose results before fluorescein injection were 7.4 mmol/l (SD 4.5) with the glucose meter and 8.1 mmol/l (SD 5.0) biochemically (p=0.02). The post dose levels were 7.2 mmol/l (SD 4.7) and 8.5 mmol/l (SD 4.8), respectively (p=0.001). The difference between the mean results of the two tests was 0.7 mmol/l before fluorescein and 1.3 mmol/l after fluorescein (p=0.001)
CONCLUSION: Intravenous fluorescein appears to influence minimally the difference between glucose levels measured in the laboratory and by glucose meter. The degree of difference, whilst achieving statistical significance, should not affect insulin dosage clinically.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10321538     DOI: 10.1034/j.1600-0420.1999.770216.x

Source DB:  PubMed          Journal:  Acta Ophthalmol Scand        ISSN: 1395-3907


  2 in total

1.  Systemic reaction to and interference with urine protein measurements after intravenous fluorescein injection.

Authors:  Annemieke Clementine Bouman; Damaris de Kruijf; Judith Maria Antonia Emmen; Thijs Clemens David Rettig; Maria Jolande Charlotte Kooijman-de Groot; Miriam Christina Faes
Journal:  BMC Ophthalmol       Date:  2019-12-19       Impact factor: 2.209

2.  Evaluation of three glucometers for whole blood glucose measurements at the point of care in preterm or low-birth-weight infants.

Authors:  Joon Ho Hwang; Yong-Hak Sohn; Seong-Sil Chang; Seung Yeon Kim
Journal:  Korean J Pediatr       Date:  2015-08-21
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.