Validation of an automated arterial tonometry monitor using Association for the Advancement of Medical Instrumentation standards.

.   

Abstract

BACKGROUND: Arterial tonometry is a noninvasive technique for monitoring the arterial blood pressure in a continuous manner. Real-time arterial blood pressure measurements are advantageous in many clinical settings, but arterial tonometric devices must be validated prior to clinical acceptance. Guidelines for accuracy for noninvasive blood pressure monitors have been established by the Association for the Advancement of Medical Instrumentation.
OBJECTIVE: To test the Colin Pilot 9200 configured with an arterial tonometry module with 20 patients and to compare tonometric blood pressure measurements with intra-arterial blood pressure measurements.
METHODS: All of the patients in the study were aged over 14 years and weighed over 35 kg; testing was performed in the operating room or in the intensive care unit. Data from each patient consisted of multiple simultaneous recordings of tonometric and intra-arterial blood pressure values. The data were then compared; the mean and SD of the difference between the two measurement devices were then calculated.
RESULTS: Tonometric values were slightly less than the intra-arterial pressure measurements; the mean difference for systolic blood pressure was 2.24 +/- 8.7 mmHg and for diastolic pressure was 0.26 +/- 8.88 mmHg.
CONCLUSION: The arterial tonometry module incorporated into the Colin Pilot 9200 was investigated for use with selected adult and pediatric populations. With our patients, it generated accurate data throughout a wide blood pressure range. It satisfied Association for the Advancement of Medical Instrumentation standards for mean systolic and diastolic blood pressure measurements and only minimally exceeded the allowable SD. This technology should prove to be a valuable tool for noninvasive blood pressure monitoring of various patient populations.

Entities:  

Year:  1997        PMID: 10234114

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  6 in total

1.  Blood pressure measuring devices: recommendations of the European Society of Hypertension.

Authors:  E O'Brien; B Waeber; G Parati; J Staessen; M G Myers
Journal:  BMJ       Date:  2001-03-03

2.  Cerebral vasoreactivity during hypercapnia is reset by augmented sympathetic influence.

Authors:  Peizhen Zhang; Guoyuan Huang; Xiangrong Shi
Journal:  J Appl Physiol (1985)       Date:  2010-11-11

3.  Cardiovagal autonomic function in sedentary and trained offspring of hypertensive parents.

Authors:  Zsuzsanna Lénárd; Péter Studinger; Beatrix Mersich; Gábor Pavlik; Mark Kollai
Journal:  J Physiol       Date:  2005-03-17       Impact factor: 5.182

Review 4.  The relationship between orthostatic hypotension and falling in older adults.

Authors:  Brett H Shaw; Victoria E Claydon
Journal:  Clin Auton Res       Date:  2013-11-20       Impact factor: 4.435

5.  Noninvasive Intracranial Pressure Assessment in Patients with Suspected Idiopathic Intracranial Hypertension.

Authors:  Bernhard Schmidt; Marek Czosnyka; Danilo Cardim; Bernhard Rosengarten
Journal:  Acta Neurochir Suppl       Date:  2021

6.  Validation of the pulse decomposition analysis algorithm using central arterial blood pressure.

Authors:  Martin C Baruch; Kambiz Kalantari; David W Gerdt; Charles M Adkins
Journal:  Biomed Eng Online       Date:  2014-07-08       Impact factor: 2.819

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.