Influence of stent length and heparin coating on platelet activation: a flow cytometric analysis in a pulsed floating model.

C Beythien1, K Gutensohn, J Bau, C W Hamm, P Kühnl, T Meinertz, W Terres.   

Abstract

Platelets are involved in acute and subacute thrombotic occlusions of coronary stents and also may play a role in the pathophysiology of in-stent restenosis. This study sought to investigate the expression of activation dependent glycoproteins on platelets by flow cytometry and time until stent thrombosis in an in vitro model of stent thrombosis. Coronary stents were placed in parallel silicon tubings with circulating citrated platelet rich plasma to measure 1) influence of stent length on platelet antigens; 2) influence of heparin coating on platelet antigens; and 3) time until stent thrombosis. After recalcification aliquots of platelet-rich plasma were taken over 10 minutes in 2-minute intervals and immediately fixed and stabilized. For flow cytometric analysis monoclonal antibodies to CD41a (glycoprotein IIb/ IIIa), CD42b (glycoprotein Ib-V-IX), CD62p (P-selectin), and CD63 (glycoprotein 53) were used. Within 2 minutes after start of circulation, the expression of CD62p and CD63 increased. Longer stents resulted in more platelet activation than shorter stents (25 mm vs. 15 mm; p<0.001. Time until stent thrombosis was reduced (25 mm vs. 15 mm; p<0.05). Heparin coating did not significantly influence flow cytometry detectable platelet activation but prolonged time until stent thrombosis (coated vs. uncoated; p<0.005). In control tubing systems without stents platelet activation was less pronounced (p<0.0001). Antibodies to CD41a and CD42b did not show significant changes. In this model platelet activation detected by flow cytometry and time until stent thrombosis were dependent on stent length and coating. In vitro testing could be useful to optimize stent design and material.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10230892     DOI: 10.1016/s0049-3848(98)00198-4

Source DB:  PubMed          Journal:  Thromb Res        ISSN: 0049-3848            Impact factor:   3.944


  4 in total

Review 1.  Coronary stents: in these days of climate change should all stents wear coats?

Authors:  R Lowe; I B A Menown; G Nogareda; I M Penn
Journal:  Heart       Date:  2005-06       Impact factor: 5.994

2.  A novel in vitro model for preclinical testing of the hemocompatibility of intravascular stents according to ISO 10993-4.

Authors:  Stefan Sinn; Torsten Scheuermann; Stephan Deichelbohrer; Gerhard Ziemer; Hans P Wendel
Journal:  J Mater Sci Mater Med       Date:  2011-05-21       Impact factor: 3.896

Review 3.  Anti-thrombogenic coatings for devices in neurointerventional surgery: Case report and review of the literature.

Authors:  Hans Henkes; Pervinder Bhogal; Marta Aguilar Pérez; Tim Lenz-Habijan; Catrin Bannewitz; Marcus Peters; Christina Sengstock; Oliver Ganslandt; Pedro Lylyk; Hermann Monstadt
Journal:  Interv Neuroradiol       Date:  2019-06-27       Impact factor: 1.610

4.  Fenestrated stent graft repair of abdominal aortic aneurysm: hemodynamic analysis of the effect of fenestrated stents on the renal arteries.

Authors:  Zhonghua Sun; Thanapong Chaichana
Journal:  Korean J Radiol       Date:  2009-12-28       Impact factor: 3.500

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.