Performance of a reagent strip device for quantitation of the urine albumin: creatinine ratio in a point of care setting.

M Parsons1, D J Newman, M Pugia, R G Newall, C P Price.   

Abstract

AIM: We have evaluated the performance of a reagent strip (Clinitek, Bayer plc, Newbury) incorporating a novel dip and read device for the quantitation of the albumin: creatinine ratio together with their individual concentrations in urine.
METHODS: The performance was compared with that of a lateral flow device for the semi-quantitation of albumin (Micral II, Roche Diagnostics, Lewes) and also a laboratory based procedure. The device employs novel methods for both analytes, using a sulphonephthalein dye binding at pH 1.5 for albumin and the peroxidase-like activity of copper creatinine complexes. The color yields of the separate reaction pads are monitored with the Clinitek 50TM bench top urine chemistry analyzer and compared to a pre-programmed calibration algorithm.
RESULTS: The imprecision of the device was assessed by observing the discrepancy between duplicates in a total of 144 urine samples collected from patients with diabetes and/or renal disease; there were 10 discrepancies in the case of the albumin estimation (6.9%), 12 in the case of the creatinine estimation (12.5%) and 23 in the case of the albumin: creatinine ratio (16.0%). In the case of the Micral II, where 96 of the urines were analyzed there were 12 discrepancies (12.5%). When considered as a two-class test for albumin with a cut off of 20 mg/L the Clinitek gave a sensitivity of 95.4% with specificity of 78.9% and a positive predictive value of 87.4%. When analyzing four urines (two controls, two patient pools) with replicate analysis on the Clinitek system we found 100% agreement for the albumin estimation, 95% for creatinine and 96.7% for the albumin: creatinine for 60 analysis within a day and 100, 95 and 97.5% on single analyses each day for 20 days for the two urine controls. The discrepant results were always within one color block. When considering the albumin: creatinine ratio with a cut off of < 30 mg/g the Clinitek gave a sensitivity of 76.3% with a specificity of 89.1% and a positive predictive value of 89.7%.
CONCLUSION: The Clinitek system provides a reliable means to screen for microalbuminuria with the opportunity of a semi-quantitative assessment when microalbuminuria is found.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10230554

Source DB:  PubMed          Journal:  Clin Nephrol        ISSN: 0301-0430            Impact factor:   0.975


  9 in total

Review 1.  Complications of diabetes: renal disease and promotion of self-management.

Authors:  A Melville; R Richardson; D Lister-Sharp; A McIntosh
Journal:  Qual Health Care       Date:  2000-12

2.  Chapter 1: Definition and classification of CKD.

Authors: 
Journal:  Kidney Int Suppl (2011)       Date:  2013-01

Review 3.  Paper-based assays for urine analysis.

Authors:  Eric Lepowsky; Fariba Ghaderinezhad; Stephanie Knowlton; Savas Tasoglu
Journal:  Biomicrofluidics       Date:  2017-10-17       Impact factor: 2.800

Review 4.  Urinary strips for protein assays: easy to do but difficult to interpret!

Authors:  Guillaume Résimont; Laurence Piéroni; Edith Bigot-Corbel; Etienne Cavalier; Pierre Delanaye
Journal:  J Nephrol       Date:  2020-04-23       Impact factor: 3.902

5.  Prevalence of Early Stages of Chronic Kidney Disease in Healthy Army Personnel.

Authors:  P P Varma; D K Raman; T S Ramakrishnan; Pragnya Singh
Journal:  Med J Armed Forces India       Date:  2011-07-21

6.  Prevalence, risk factors and awareness of albuminuria on a Canadian First Nation: a community-based screening study.

Authors:  James Michael Zacharias; T Kue Young; Natalie D Riediger; Joanne Roulette; Sharon G Bruce
Journal:  BMC Public Health       Date:  2012-04-20       Impact factor: 3.295

7.  Diagnostic accuracy of urine dipstick testing for albumin-to-creatinine ratio and albuminuria: A systematic review and meta-analysis.

Authors:  Jhonatan R Mejia; Jose Ernesto Fernandez-Chinguel; Gandy Dolores-Maldonado; Naysha Becerra-Chauca; Sergio Goicochea-Lugo; Percy Herrera-Añazco; Jessica Hanae Zafra-Tanaka; Alvaro Taype-Rondan
Journal:  Heliyon       Date:  2021-10-23

8.  Microalbuminuria measured by three different methods, blood pressure and cardiovascular risk factors in elderly Swedish males.

Authors:  Gösta Florvall; Samar Basu; Johanna Helmersson; Anders Larsson
Journal:  Anal Chem Insights       Date:  2008-09-02

Review 9.  Comparison of commonly used assays for the detection of microalbuminuria.

Authors:  Douglas E Busby; George L Bakris
Journal:  J Clin Hypertens (Greenwich)       Date:  2004-11       Impact factor: 3.738

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.