| Literature DB >> 10226131 |
A Rossetti1, M Paccosi, O Sizzi, S Zulli, S Mancuso, A Lanzone.
Abstract
To improve enucleation of uterine myomas, we designed a reusable 5-mm laparoscopic drill with a distal forked pin and locking system. Because the device is jointed, it can apply traction not only along its own axis, but also along planes, depending on its bending radius. We compared enucleation time for myomas 5 to 8 cm in size before and after introduction of the drill. The last 23 procedures performed with this instrument were easier, with a reduction in operating time of about 28.5% for the enucleation part of the procedure (p <0.001). (J Am Assoc Gynecol Laparosc 6(2):189-193, 1999)Entities:
Mesh:
Substances:
Year: 1999 PMID: 10226131 DOI: 10.1016/s1074-3804(99)80101-7
Source DB: PubMed Journal: J Am Assoc Gynecol Laparosc ISSN: 1074-3804