Literature DB >> 10225151

Stapedotomy: does prosthesis diameter affect outcome?

Y K Shabana1, M R Ghonim, C B Pedersen.   

Abstract

Two hundred ears with otosclerosis have been operated upon using a 0.4-mm diameter Teflon platinum piston in 100 ears and a 0.6-mm diameter Teflon piston in the other 100 ears. The postoperative air-bone gap, calculated as the difference between the postoperative air and bone conduction levels, was smaller in the 0.6 mm group for all frequencies except at 2000 Hz, the differences were statistically insignificant except at 4000 Hz. The mean postoperative air-bone gap was 8.6 dB and 7.4 dB for the 0.4 and 0.6 mm groups, respectively, which is statistically insignificant. We found no postoperative loss of bone conduction exceeding 15 dB, there was a deterioration of more than 10 dB in three ears, one in the 0.4 mm group and two in the 0.6 mm group. According to our results, we conclude that the 0.4 mm and the 0.6 mm Teflon prostheses produce the same hearing improvement in stapes surgery for otosclerosis.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10225151     DOI: 10.1046/j.1365-2273.1999.00207.x

Source DB:  PubMed          Journal:  Clin Otolaryngol Allied Sci        ISSN: 0307-7772


  3 in total

1.  Stapes surgery: a National Survey of British Otologists.

Authors:  Hannah Lancer; Jaiganesh Manickavasagam; Azreena Zaman; Jack Lancer
Journal:  Eur Arch Otorhinolaryngol       Date:  2015-02-25       Impact factor: 2.503

Review 2.  Design and optimization of auditory prostheses using the finite element method: a narrative review.

Authors:  Qianli Cheng; Han Yu; Junpei Liu; Qi Zheng; Yanru Bai; Guangjian Ni
Journal:  Ann Transl Med       Date:  2022-06

3.  Stimulation efficiency of an actuator driven piston at the biological interface to the inner ear.

Authors:  Susan Busch; Mohammad Ghoncheh; Thomas Lenarz; Hannes Maier
Journal:  Sci Rep       Date:  2021-12-09       Impact factor: 4.379

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.