Literature DB >> 10221371

Comparison of two home blood pressure monitors with a mercury sphygmomanometer in an ambulatory population.

K A Johnson1, D J Partsch, P Gleason, K Makay.   

Abstract

We performed a randomized, single-visit, crossover study to compare the precision of a digital monitor and an aneroid monitor relative to a desktop mercury sphygmomanometer (DMS). Three blood pressure measurements per patient, one per device, were taken over 25 minutes. Of 99 patients, 95 had complete blood pressure data. Systolic blood pressures measured with the digital monitor were significantly different from those with the DMS (p<0.01). Thirty-four percent of systolic and 48% of diastolic pressures measured with the digital monitor were within +/- 5 mm Hg of the DMS. The aneroid monitor performed significantly better than the digital monitor, with 54% of systolic (p<0.01) and 58% of diastolic blood pressures within +/- 5 mm Hg of the DMS standard. Health care providers are encouraged to supervise patients' initial use of a home blood pressure monitor and calibrate readings against a standard mercury sphygmomanometer.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10221371     DOI: 10.1592/phco.19.4.333.30936

Source DB:  PubMed          Journal:  Pharmacotherapy        ISSN: 0277-0008            Impact factor:   4.705


  1 in total

1.  Which is More Accurate in Measuring the Blood Pressure? A Digital or an Aneroid Sphygmomanometer.

Authors:  Bhaskar Shahbabu; Aparajita Dasgupta; Kaushik Sarkar; Sanjaya Kumar Sahoo
Journal:  J Clin Diagn Res       Date:  2016-03-01