Literature DB >> 10201755

Evaluation of endoscope sheaths as viral barriers.

K H Baker1, M P Chaput, C R Clavet, G W Varney, T M To, C D Lytle.   

Abstract

OBJECTIVES: Evaluate ENT endoscope sheaths as barriers to virus passage. STUDY
DESIGN: "Defective" sheaths covering an endoscope were challenged with virus to determine how many virus particles could be recovered from the endoscope.
METHODS: Sheaths with small laser-drilled holes (2 to 30 microm) were challenged with high-titer virus suspensions (10(8) viruses/mL). The inside of the sheath and the endoscope were separately rinsed to recover any virus that penetrated through the hole in the sheath. In an attempt to assess the possible importance of holes in the sheaths, a sequential test was conducted with an initial virus challenge outside a defective sheath (30-micron hole in the sheath), after which the possibly contaminated endoscope was removed and inserted into a second defective sheath (with a 20-micron hole at the same location) to determine whether the contaminating virus would pass outward through the second sheath.
RESULTS: Small volumes of virus-containing fluid penetrated through the hole, e.g., 500 virus particles passed through one of three 30-microm holes. A significant fraction of those virus particles was occasionally found on the endoscope after removal from the sheath. Similar results were obtained with sheaths that had small tears (34-84 microm in length, from punctures with fine wires). Although some virus penetration could occur during the initial challenge contaminating the endoscope, no virus was detected passing outward through the second sheath.
CONCLUSIONS: Use of a sheath combined with intermediate level disinfection should provide a safe instrument for ENT endoscopy.

Mesh:

Year:  1999        PMID: 10201755     DOI: 10.1097/00005537-199904000-00022

Source DB:  PubMed          Journal:  Laryngoscope        ISSN: 0023-852X            Impact factor:   3.325


  2 in total

1.  Evaluating the Quality of Rigid Optic Videolaryngoscopy Image Taken Through Dental Protection Cap and Its Feasibility as Additional Barrier Method Against COVID-19.

Authors:  Mariana Nagata Cavalheiro; Vinicius Ribas Fonseca; Bianca Simone Zeigelboim; Diego F Costa; Lucas Viomar de Lima; Marlon Kleber Bozzo; Rita de Cássia Tonochi; Bruno Ceron Hartmann
Journal:  J Voice       Date:  2020-11-02       Impact factor: 2.009

2.  A microbiological evaluation of level of disinfection for flexible cystoscopes protected by disposable endosheaths.

Authors:  Peter Hjorth Jørgensen; Torsten Slotsbjerg; Henrik Westh; Vicki Buitenhuis; Gregers Gautier Hermann
Journal:  BMC Urol       Date:  2013-10-07       Impact factor: 2.264

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.