Mapping of atrial activation with a noncontact, multielectrode catheter in dogs.

A Kadish1, J Hauck, B Pederson, G Beatty, C Gornick.   

Abstract

BACKGROUND: Endocardial mapping of sustained arrhythmias has traditionally been performed with a roving diagnostic catheter. Although this approach is adequate for many tachyarrhythmias, it has limitations. The purpose of this study was to evaluate a novel noncontact mapping system for assessing atrial tachyarrhythmias. METHODS AND
RESULTS: The mapping system consists of a 9F multielectrode-array balloon catheter that has 64 active electrodes and ring electrodes for emitting a locator signal. The locator signal was used to construct a 3-dimensional right atrial map; it was independently validated and was highly accurate. Virtual electrograms were calculated at 3360 endocardial sites in the right atrium. We evaluated right atrial activation by positioning the balloon catheter in the mid right atrium via a femoral venous approach. Experiments were performed on 12 normal mongrel dogs. The mean correlation coefficient between contact and virtual electrograms was 0.80+/-0.12 during sinus rhythm. Fifty episodes of atrial flutter induced in 11 animals were evaluated. In the majority of experiments, complete or almost complete reentrant circuits could be identified within the right atrium. Mean correlation coefficient between virtual and contact electrograms was 0.85+/-0.17 in atrial flutter. One hundred fifty-six episodes of pacing-induced atrial fibrillation were evaluated in 11 animals. Several distinct patterns of right atrial activation were seen, including single-activation wave fronts and multiple simultaneous-activation wave fronts. Mean correlation coefficient between virtual and contact electrograms during atrial fibrillation was 0.81+/-0.18. The accuracy of electrogram reconstruction was lower at sites >4.0 cm from the balloon center and at sites with a high spatial complexity of electrical activation.
CONCLUSIONS: This novel noncontact mapping system can evaluate conduction patterns during sinus rhythm, demonstrate reentry during atrial flutter, and describe right atrial activation during atrial fibrillation. The accuracy of electrogram reconstruction was good at sites <4.0 cm from the balloon center, and thus the system has the ability to perform high-resolution multisite mapping of atrial tachyarrhythmias in vivo.

Entities:  

Mesh:

Year:  1999        PMID: 10199890     DOI: 10.1161/01.cir.99.14.1906

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  21 in total

Review 1.  Telomerase-independent mechanisms of telomere elongation.

Authors:  H Biessmann; J M Mason
Journal:  Cell Mol Life Sci       Date:  2003-11       Impact factor: 9.261

2.  Non-contact left ventricular endocardial mapping in cardiac resynchronisation therapy.

Authors:  P D Lambiase; A Rinaldi; J Hauck; M Mobb; D Elliott; S Mohammad; J S Gill; C A Bucknall
Journal:  Heart       Date:  2004-01       Impact factor: 5.994

3.  Radiofrequency catheter ablation using non-contact mapping for inappropriate sinus tachycardia.

Authors:  Chad E Bonhomme; Florin T Deger; Jeff Schultz; Steve S Hsu
Journal:  J Interv Card Electrophysiol       Date:  2004-04       Impact factor: 1.900

Review 4.  Cell delivery and tracking in post-myocardial infarction cardiac stem cell therapy: an introduction for clinical researchers.

Authors:  Heming Wei; Ting Huay Ooi; Genevieve Tan; Sze Yun Lim; Ling Qian; Philip Wong; Winston Shim
Journal:  Heart Fail Rev       Date:  2010-01       Impact factor: 4.214

Review 5.  Rotors as drivers of atrial fibrillation and targets for ablation.

Authors:  Amir A Schricker; Gautam G Lalani; David E Krummen; Sanjiv M Narayan
Journal:  Curr Cardiol Rep       Date:  2014-08       Impact factor: 2.931

6.  High-resolution noncontact charge-density mapping of endocardial activation.

Authors:  Andrew Grace; Stephan Willems; Christian Meyer; Atul Verma; Patrick Heck; Min Zhu; Xinwei Shi; Derrick Chou; Lam Dang; Christoph Scharf; Günter Scharf; Graydon Beatty
Journal:  JCI Insight       Date:  2019-03-21

7.  Mechanisms of ventricular fibrillation in canine models of congestive heart failure and ischemia assessed by in vivo noncontact mapping.

Authors:  Thomas H Everett; Emily E Wilson; Scott Foreman; Jeffrey E Olgin
Journal:  Circulation       Date:  2005-09-06       Impact factor: 29.690

8.  Transmural characteristics of atrial fibrillation in canine models of structural and electrical atrial remodeling assessed by simultaneous epicardial and endocardial mapping.

Authors:  Thomas H Everett; Emily E Wilson; George S Hulley; Jeffrey E Olgin
Journal:  Heart Rhythm       Date:  2010-01-11       Impact factor: 6.343

9.  Transseptal activation during left atrial pacing in humans: electroanatomic mapping using a noncontact catheter and the intracardiac echocardiography.

Authors:  Leonardo Calò; Filippo Lamberti; Maria Luisa Loricchio; Antonio Castro; Augusto Boggi; Furio Colivicchi; Claudio Pandozi; Massimo Santini
Journal:  J Interv Card Electrophysiol       Date:  2002-06       Impact factor: 1.900

Review 10.  Catheter ablation of ventricular tachycardia. From indication to three-dimensional mapping technology.

Authors:  Lars Eckardt; Günter Breithardt
Journal:  Herz       Date:  2009-05       Impact factor: 1.443

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.