Literature DB >> 10198720

A preliminary study of microcapsule suspension for hemolysis evaluation of artificial organs.

O Maruyama1, T Yamane, N Tsunemoto, M Nishida, T Tsutsui, T Jikuya.   

Abstract

A microcapsule suspension, a substitute for animal blood in hemolysis tests, has been developed for evaluation of the absolute hemolytic properties of circulatory artificial organs. The microcapsule suspension was made by dispersing microcapsule slurry into an ethylene glycol sodium chloride solution. The microcapsule slurry was composed of a leuco dye solution and polyurethane membrane made by the reaction between aliphatic poly-isocyanate and polyamine by interfacial polycondensation. The microcapsule was a small particle containing dye inside. The microcapsule suspension was white; the diameter of the microcapsules was from 5 to 100 microns. The specific gravity of the suspension was 1.024, and the membrane was elastic. The fluid showed Newtonian characteristics, different from animal blood, and its viscosity was approximately 5.8 mPa.s. After the microcapsules were destroyed, the leuco dye was extracted with n-hexane from the suspension and was measured by spectroscopy after being colored with acid ethanol. Hemolysis can be regarded as a fatigue fracture of cell membranes rather than a static fracture. The destruction of microcapsules by a Potter type tissue grinder was observed at a low stroke number region and was compared to rat blood. Moreover, hemolysis tests of a commercially available centrifugal blood pump and the prototype of our centrifugal pump for mechanism checks were carried out with bovine blood. The hemolysis level of the prototype pump increased with time while the hemolysis level of the commercial blood pump did not change as much as that of the control when both pumps were tested with the microcapsule suspension. These results are similar to tests utilizing bovine blood. Therefore, hemolysis tests of circulatory artificial organs completed with microcapsule suspension are expected to provide results similar to tests with animal blood.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10198720     DOI: 10.1046/j.1525-1594.1999.06316.x

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


  3 in total

1.  In vitro particulate analogue fluids for experimental studies of rheological and hemorheological behavior of glucose-rich RBC suspensions.

Authors:  Diana Pinho; Laura Campo-Deaño; Rui Lima; Fernando T Pinho
Journal:  Biomicrofluidics       Date:  2017-09-21       Impact factor: 2.800

2.  Particulate Blood Analogues Reproducing the Erythrocytes Cell-Free Layer in a Microfluidic Device Containing a Hyperbolic Contraction.

Authors:  Joana Calejo; Diana Pinho; Francisco J Galindo-Rosales; Rui Lima; Laura Campo-Deaño
Journal:  Micromachines (Basel)       Date:  2015-12-30       Impact factor: 2.891

3.  Microfluidic Deformability Study of an Innovative Blood Analogue Fluid Based on Giant Unilamellar Vesicles.

Authors:  Denise A M Carvalho; Ana Rita O Rodrigues; Vera Faustino; Diana Pinho; Elisabete M S Castanheira; Rui Lima
Journal:  J Funct Biomater       Date:  2018-12-04
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.