| Literature DB >> 10160448 |
E J Snyder1, D L McElwee, H M Harb, J A Cullen, I S Tackel, S Baumgart, B Iskandarani, A Elkind, M Shaker.
Abstract
A failure analysis was performed on Norton S-65-HL Tygon tubing. Fatigue testing was performed on four sizes of this tubing, and essentially showed how the tubing wears out. A dynamic "life hours to failure" test, which was performed on the 3/8" internal diameter (ID) size, quantified when the tubing ruptured. Based on results of laboratory testing and the institution's clinical extracorporeal membrane oxygenation (ECMO) experience, a reasonable life expectancy for the 3/8" S-65-HL Tygon size was determined for use in this institution's neonatal ECMO system. An understanding of the expected performance of roller pump tubing-an integral component of the ECMO system-is imperative to providing safe, effective extracorporeal life support.Entities:
Mesh:
Year: 1996 PMID: 10160448
Source DB: PubMed Journal: J Extra Corpor Technol ISSN: 0022-1058