C A Brandon1. 1. Department of Microbiology and Immunology, University of North Dakota, Grand Forks, USA.
Abstract
OBJECTIVE: To compare the results of urinalysis screenings for specific gravity (SG) using the reagent strip and the refractometer. SETTING: United Hospital, Grand Forks, North Dakota. PRACTICE DESCRIPTION: United Hospital is a 384-bed teaching hospital. PRODUCT COMPARISON: The Ames Multistix 10 SG reagent strip (Miles, Inc., Elkhart, IN 46515) was compared with the TS Meter (Leica, Inc., Deerfield, IL 60015). MAIN OUTCOME MEASUREMENT: The degree of correlation between the results produced by each method. RESULTS: The percentage of difference between the means of the direct strip readings and the refractometer readings was 9.68%. The percentage of difference between the means of the adjusted strip readings and the refractometer readings was 22.58%, which was significantly different. When the direct strip readings and the refractometer readings were plotted together on a graph, the points were widely scattered; this fact, and a correlation coefficient of 0.725, suggest that random error occurred in both methods. Analysis of the slope and intercept of the correlation indicated systematic error. CONCLUSION: The reagent strip method of measuring SG is accurate only in a narrow range of "average" values, and should not be used as the basis for medical diagnoses.
OBJECTIVE: To compare the results of urinalysis screenings for specific gravity (SG) using the reagent strip and the refractometer. SETTING: United Hospital, Grand Forks, North Dakota. PRACTICE DESCRIPTION: United Hospital is a 384-bed teaching hospital. PRODUCT COMPARISON: The Ames Multistix 10 SG reagent strip (Miles, Inc., Elkhart, IN 46515) was compared with the TS Meter (Leica, Inc., Deerfield, IL 60015). MAIN OUTCOME MEASUREMENT: The degree of correlation between the results produced by each method. RESULTS: The percentage of difference between the means of the direct strip readings and the refractometer readings was 9.68%. The percentage of difference between the means of the adjusted strip readings and the refractometer readings was 22.58%, which was significantly different. When the direct strip readings and the refractometer readings were plotted together on a graph, the points were widely scattered; this fact, and a correlation coefficient of 0.725, suggest that random error occurred in both methods. Analysis of the slope and intercept of the correlation indicated systematic error. CONCLUSION: The reagent strip method of measuring SG is accurate only in a narrow range of "average" values, and should not be used as the basis for medical diagnoses.