Abstract

Patients requiring extracorporeal membrane oxygenation (ECMO) often become totally dependent on the mechanical life support. The Extracorporeal Life Support Organization (ELSO) reports 2486 incidents of mechanical complications in 5905 ECMO supports. To help decrease the number of mechanical complications, an active quality assurance program was initiated at our institution. This resulted in identification of only 14 incidents of mechanical complications in 100 patients (neonate, pediatric, adult, and cardiac). Techniques for dealing with problems such as loss of roller pump occlusion, changing out of the membrane lung or heat exchanger without interrupting ECMO support, venous air lock, tamponade, emergency transfusion, and other situations were generated into written policies and procedures. We routinely review and practice problem solving techniques with specific emphasis on monitoring patient hemodynamics and appearance. We conclude that written policies and procedures, "water drills," and continuing education can be beneficial in early recognition, intervention, and/or prevention of ECMO mechanical complications.

Entities:  

Mesh:

Year:  1993        PMID: 10148324

Source DB:  PubMed          Journal:  J Extra Corpor Technol        ISSN: 0022-1058


  2 in total

1.  Simulation versus live tissue training randomised trial for ECMO proficiency: is one better than the other?

Authors:  Thornton Mu; Tricia Garcia-Choudary; Amanda Staudt; Melissa Tyree; Krystal Valdez-Delgado; Nicole Caldwell; Nicholas Carr; Matthew Borgman; Heather Delaney
Journal:  BMJ Simul Technol Enhanc Learn       Date:  2020-09-04

2.  Dermabond as a Novel Solution for Needle-Hole Repair in Arterial Cannula while on Extracorporeal Membrane Oxygenation Therapy.

Authors:  Janani S Reisenauer; James R Neal; David L Joyce
Journal:  J Extra Corpor Technol       Date:  2019-09
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.